Reproducibility and predictive value of a simple novel method to measure atrial fibrillation cycle length in persistent atrial fibrillation - FARS-AF study
- PMID: 35132713
- DOI: 10.1111/jce.15401
Reproducibility and predictive value of a simple novel method to measure atrial fibrillation cycle length in persistent atrial fibrillation - FARS-AF study
Abstract
Introduction: Different methods are used for atrial fibrillation (AF) cycle length (CL) measurement with variable results. Previous studies of pulmonary vein (PV) CL measurement showed contradictory results on predicting PV isolation (PVI) efficacy. A novel simple method of measuring the average of 10 consecutive Fastest Atrial Repetitive Similar morphology signal (FARS10 )-CL to characterize local atrial activity rate was evaluated prospectively.
Methods: The intra-observer reproducibility of FARS10 -CL and traditional AF-CL measurement of continuously fragmented coronary sinus (CS) signals were tested. We prospectively enrolled 100 consecutive patients (62 ± 10 years, 72% male) undergoing wide antral PVI only ablation for persistent AF, measured PV-FARS10 -CLs, and evaluated long-term outcome.
Results: The Kendall area correlation between repeated traditional AF-CL measurements was -0.006 and between repeated FARS10 -CL measurements in the right and left atrial appendages, CS and PVs were 0.944, 0.859, 0.882, 0.675-0.955, respectively. Patients with recurrent atrial tachyarrhythmia had significantly longer fastest PV-FARS10 -CL (172 ± 41 vs. 156 ± 41 ms, p = .047). Patients with high burden of spontaneous low-voltage zone (LVZ) had significantly longer fastest PV-FARS10 -CL. Freedom from recurrent tachyarrhythmia at 24 months was 85% versus 59% in patients with fastest PV-FARS10 -CL ≤ 140 versus >140 ms, p = .0018, respectively. In multivariable analysis fastest PV-FARS10 -CL ≤ 140 ms was the only significant predictor of freedom from recurrent tachyarrhythmia.
Conclusions: FARS10 -CL measurements have a high reproducibility in contrast to traditional AF-CL measurement of continuously fragmented CS signals. Patients with high burden of LVZ have longer fastest PV-FARS10 -CLs. Fastest PV-FARS10 -CL ≤ 140 ms is associated with a high success of wide antral PVI-only ablation approach in persistent AF.
Keywords: atrial fibrillation ablation; atrial fibrillation cycle length; persistent atrial fibrillation; predictors AF ablation; pulmonary vein isolation.
© 2022 Wiley Periodicals LLC.
Comment in
-
The Fa(r)st or the spurious - pulmonary vein cycle length: Is it the fuel driving persistent AF, or simply just fumes?J Cardiovasc Electrophysiol. 2022 Apr;33(4):651-653. doi: 10.1111/jce.15406. Epub 2022 Feb 15. J Cardiovasc Electrophysiol. 2022. PMID: 35132726 No abstract available.
References
REFERENCES
-
- Calkins H, Hindricks G, Cappato R, et al. 2017 HRS/EHRA/ECAS/APHRS/SOLAECE expert consensus statement on catheter and surgical ablation of atrial fibrillation: executive summary. Europace. 2018;20(1):e1-e160.
-
- O′Donnell D, Furniss SS, Bourke JP. Paroxysmal cycle length shortening in the pulmonary veins during atrial fibrillation correlates with arrhythmogenic triggering foci in sinus rhythm. J Cardiovasc Electrophysiol. 2002;13:124-128.
-
- Verma A, Jiang CY, Betts TR, et al. Approaches to catheter ablation for persistent atrial fibrillation. N Engl J Med. 2015;372(19):1812-1822.
-
- Chen SA, Tai CT. Catheter ablation of atrial fibrillation originating from the non-pulmonary vein foci. J Cardiovasc Electrophysiol. 2005;16(2):229-232.
-
- Santangeli P, Zado ES, Hutchinson MD, et al. Prevalence and distribution of focal triggers in persistent and long-standing persistent atrial fibrillation. Heart Rhythm. 2016;13(2):374-382.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials