Computational Analysis of the Droplet-Stimulated Laryngeal Adductor Reflex in High-Speed Sequences
- PMID: 35133015
- DOI: 10.1002/lary.30041
Computational Analysis of the Droplet-Stimulated Laryngeal Adductor Reflex in High-Speed Sequences
Abstract
Objectives/hypothesis: The laryngeal adductor reflex (LAR) is an important protective mechanism of the airways. Its physiology is still not completely understood. The available methods for LAR evaluation offer limited reproducibility and/or rely on subjective interpretation. A new approach, termed Microdroplet Impulse Testing of the LAR (MIT-LAR), was recently introduced. Here, the LAR is elicited by a droplet and a laryngoscopic high-speed recording is acquired simultaneously. In the present work, image-processing algorithms for autonomous MIT-LAR sequence analysis were developed. This allowed the automated approximation of kinematic LAR parameters in humans.
Study design: Development and testing of computational methods.
Methods: Computational image processing enabled the autonomous estimation of the glottal area, the glottal angle, and the vocal fold edge distance in MIT-LAR sequences. A suitable analytical representation of these glottal parameters allowed the extraction of seven relevant LAR parameters. The obtained values were compared to the literature.
Results: A generalized logistic function showed the highest average goodness of fit among four different analytical approaches for each of the glottal parameters. Autonomous sequence analysis yielded bilateral LAR response latencies of (229 ± 116) ms and (182 ± 60) ms for cases of complete and incomplete glottal closure, respectively. The initial/average/maximum angular vocal fold adduction velocity was estimated at (157 ± 115) °s-1 /(891 ± 516) °s-1 /(929 ± 583) °s-1 and (88 ± 53) °s-1 /(421 ± 221) °s-1 /(520 ± 238) °s-1 for complete and incomplete glottal closure, respectively.
Conclusion: The automated extraction of LAR parameters from laryngoscopic high-speed sequences can potentially increase the objectiveness of optical LAR characterization and reduce the associated workload. The proposed methods may thus be helpful for future research on this vital reflex.
Level of evidence: NA Laryngoscope, 132:2412-2419, 2022.
Keywords: Airway-protective mechanisms; MIT-LAR; computer-aided diagnosis; glottal closure reflex; high-speed endoscopy; laryngeal closure reflex; laryngeal kinematics; reflex evaluation.
© 2022 The Authors. The Laryngoscope published by Wiley Periodicals LLC on behalf of The American Laryngological, Rhinological and Otological Society, Inc.
References
REFERENCES
-
- Jadcherla SR, Hogan WJ, Shaker R. Physiology and pathophysiology of glottic reflexes and pulmonary aspiration: from neonates to adults. Semin Respir Crit Care Med 2010;31:554-560. https://doi.org/10.1055/s-0030-1265896.
-
- Yamashita T, Nash EA, Tanaka Y, Ludlow CL. Effects of stimulus intensity on laryngeal long latency responses in awake humans. Otolaryngol Head Neck Surg 1997;117:521-529.
-
- Andreatta RD, Mann EA, Poletto CJ, Ludlow CL. Mucosal afferents mediate laryngeal adductor responses in the cat. J Appl Physiol 2002;93:1622-1629. https://doi.org/10.1152/japplphysiol.00417.2002.
-
- Shock LA, Gallemore BC, Hinkel CJ, et al. Improving the utility of laryngeal adductor reflex testing: a translational tale of mice and men. Otolaryngol Head Neck Surg 2015;153:94-101. https://doi.org/10.1177/0194599815578103.
-
- Strohl MP, Young VN, Dwyer CD, et al. Novel adaptation of a validated tactile aesthesiometer to evaluate laryngopharyngeal sensation. Laryngoscope 2021;131:1324-1331. https://doi.org/10.1002/lary.28947.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials
Miscellaneous