Improved accuracy and precision with three-parameter simultaneous myocardial T1 and T2 mapping using multiparametric SASHA
- PMID: 35133018
- DOI: 10.1002/mrm.29170
Improved accuracy and precision with three-parameter simultaneous myocardial T1 and T2 mapping using multiparametric SASHA
Abstract
Purpose: To develop and validate a three-parameter model for improved precision multiparametric SAturation-recovery single-SHot Acquisition (mSASHA) cardiac T1 and T2 mapping with high accuracy in a single breath-hold.
Methods: The mSASHA acquisition consists of nine images of variable saturation recovery and T2 preparation in 11 heartbeats with T1 and T2 values calculated using a three-parameter model. It was validated in simulations and phantoms at 3 T with comparison to a four-parameter joint T1 -T2 technique. The mSASHA acquisition was compared with MOLLI, SASHA, and T2 -prepared balanced SSFP in 10 volunteers.
Results: The mSASHA technique had high accuracy in phantoms compared to spin echo, with -0.2 ± 0.3% T1 error and -2.4 ± 1.3% T2 error. The mSASHA coefficient of variation in phantoms for T1 was similar to MOLLI (0.7 ± 0.2% for both) and T2 -prepared balanced SSFP for T2 (1.3 ± 0.7% vs 1.4 ± 0.3%, adjusted p > .05 for both). In simulations, three-parameter mSASHA had higher precision than four-parameter joint T1 -T2 for both T1 and T2 (46% and 11% reductions in T1 and T2 interquartile range for native myocardium). In vivo myocardial mSASHA T1 was similar to SASHA (1523 ± 18 ms vs 1520 ± 18 ms) with similar coefficient of variation to both MOLLI and SASHA (3.3 ± 0.6% vs 3.1 ± 0.6% and 3.3 ± 0.5% respectively, adjusted p > .05 for all). Myocardial mSASHA T2 was 37.1 ± 1.1 ms with similar precision to T2 -prepared balanced SSFP (6.7 ± 1.7% vs 6.0 ± 1.6%, adjusted p > .05).
Conclusion: Three-parameter mSASHA provides high-accuracy cardiac T1 and T2 quantification in a single breath-hold with similar precision to MOLLI and T2 -prepared balanced SSFP. Further study is required to both establish normative values and demonstrate clinical utility in patient populations.
Keywords: MOLLI; SASHA; T1 mapping; T2 mapping; cardiac MR; multiparametric mapping.
© 2022 International Society for Magnetic Resonance in Medicine. This article has been contributed to by US Government employees and their work is in the public domain in the USA.
References
REFERENCES
-
- Leiner T, Bogaert J, Friedrich MG, et al. SCMR Position Paper (2020) on clinical indications for cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2020;22:1-37.
-
- Bull S, White SK, Piechnik SK, et al. Human non-contrast T1 values and correlation with histology in diffuse fibrosis. Heart. 2013;99:932-937.
-
- Flacke SJ, Fischer SE, Lorenz CH. Measurement of the gadopentetate dimeglumine partition coefficient in human myocardium in vivo: normal distribution and elevation in acute and chronic infarction. Radiology. 2001;218:703-710.
-
- Ugander M, Oki AJ, Hsu L-Y, et al. Extracellular volume imaging by magnetic resonance imaging provides insights into overt and sub-clinical myocardial pathology. Eur Heart J. 2012;33:1268-1278.
-
- Sado DM, Flett AS, Banypersad SM, et al. Cardiovascular magnetic resonance measurement of myocardial extracellular volume in health and disease. Heart. 2012;98:1436-1441.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical