Background: Postimplantation syndrome (PIS) represents an acute phase systemic inflammatory response following endovascular aortic aneurysm repair (EVAR). Our objective was to investigate the risk factors associated with the manifestation and severity of PIS with various available stent-grafts.
Methods: We performed a retrospective analysis of prospectively collected data covering the period 2016-2020. The study included 191 patients. Body temperature was recorded regularly and blood sample was obtained daily. The imaging protocol included computed tomography aortoiliac angiography before surgery and one month after. The volumes of pre-existing and new-onset mural thrombus were calculated in a semi-automated fashion. Five abdominal aortic stent-graft devices were used: Endurant™ ΙΙ, Anaconda™, Treo®, E-tegra® and AFX® 2. Subgroup analysis was performed between woven polyester and ePTFE lined devices.
Results: The incidence of PIS was 21.5%. No significant differences were observed regarding demographics, risk factors, aneurysm anatomy or operative data. The amount of pre-existing and new-onset mural thrombus were not related with PIS (P=0.117 and P=0.096). PIS incidence in the polyester subgroup was 24.2%, significantly higher compared to 8.3% in the ePTFE subgroup. In-subgroup analysis revealed that the use of Anaconda™ was associated with the higher frequency (61.1%, P=0.021). Multivariate logistic regression showed that polyester was the single factor significantly associated with PIS (hazard ratio=2.6, P=0.043), as opposed to the new onset thrombus (hazard ratio=1.29, P=0.101).
Conclusions: PIS is not uncommon and should be taken into consideration in patients presenting with fever after EVAR. The endograft's liner material seems to play the primordial role, with woven polyester to be attributed with significantly higher incidence.
Similar articles
-
Post-implantation syndrome: the impact of different devices for endovascular abdominal aortic aneurysm repair and related etiopathogenetic implications.Int Angiol. 2020 Oct;39(5):398-404. doi: 10.23736/S0392-9590.20.04163-2. Epub 2020 May 13. Int Angiol. 2020. PMID: 32401471
-
Incidence of Post-Implantation Syndrome with Different Endovascular Aortic Aneurysm Repair Modalities and Devices and Related Etiopathogenetic Implications.Ann Vasc Surg. 2020 Feb;63:155-161. doi: 10.1016/j.avsg.2019.08.095. Epub 2019 Oct 16. Ann Vasc Surg. 2020. PMID: 31629127
-
Stent graft composition plays a material role in the postimplantation syndrome.J Vasc Surg. 2012 Dec;56(6):1503-9. doi: 10.1016/j.jvs.2012.06.072. Epub 2012 Oct 23. J Vasc Surg. 2012. PMID: 23092643
-
Systematic review of risk factors and outcomes of post-implantation syndrome following endovascular aortic repair.J Vasc Surg. 2024 May;79(5):1240-1250.e4. doi: 10.1016/j.jvs.2023.12.029. Epub 2023 Dec 18. J Vasc Surg. 2024. PMID: 38122858
-
Postimplantation syndrome in endovascular aortic aneurysm repair - a systematic review.Vasa. 2021 Apr;50(3):174-185. doi: 10.1024/0301-1526/a000913. Epub 2020 Nov 3. Vasa. 2021. PMID: 33138736
Cited by
-
Early Results of Elective Endovascular Repair of Infrarenal Abdominal Aortic Aneurysms With the MinosTM Stent-Graft System.J Endovasc Ther. 2025 Feb;32(1):225-232. doi: 10.1177/15266028231172379. Epub 2023 May 11. J Endovasc Ther. 2025. PMID: 37166154 Free PMC article.
-
The Long-Term Progression of Aneurysmal Disease in Common Iliac Arteries After Standard EVAR and Its Clinical Implications.Int J Vasc Med. 2024 Nov 29;2024:4229582. doi: 10.1155/ijvm/4229582. eCollection 2024. Int J Vasc Med. 2024. PMID: 39650203 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical