Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 1986 Feb;23(2):251-7.
doi: 10.1128/jcm.23.2.251-257.1986.

Critical evaluation of the AutoMicrobic system gram-negative identification card for identification of glucose-nonfermenting gram-negative rods

Critical evaluation of the AutoMicrobic system gram-negative identification card for identification of glucose-nonfermenting gram-negative rods

J J Plorde et al. J Clin Microbiol. 1986 Feb.

Abstract

During a 6-month study we critically evaluated the accuracy of the AutoMicrobic system Gram-Negative Identification Card (Vitek Systems, Inc., Hazelwood, Mo.) in identifying glucose-nonfermenting gram-negative bacilli by testing 419 selected isolates in parallel with a conventional reference method. Of 356 isolates included in the AutoMicrobic system profile, a total of 307 (86.2%) were correctly identified, 36 (10.1%) were not identified, and 13 (3.7%) were misidentified. Fifty-eight of 63 (92%) isolates not included in the profile were correctly reported as "unidentified organisms." Overall, if the first-choice identification was always accepted, only 18 (4.3%) isolates would have been incorrectly reported. When first-choice identifications appended with the special message "questionable biopattern" were rejected, and organisms were screened for characteristic odor and antimicrobial susceptibility before final acceptance of the AutoMicrobic system report, the number of misidentifications was reduced to 5 (1.2%). The average time to identification with the AutoMicrobic system Gram-Negative Identification Card was 15 h. This compares favorably with the 65 h required by the reference method.

PubMed Disclaimer

References

    1. Ann Intern Med. 1972 Aug;77(2):211-5 - PubMed
    1. Rev Infect Dis. 1984 Sep-Oct;6 Suppl 3:S627-42 - PubMed
    1. J Clin Microbiol. 1977 Dec;6(6):605-9 - PubMed
    1. Am J Clin Pathol. 1978 Jan;69(1):41-7 - PubMed
    1. J Clin Microbiol. 1978 Jun;7(6):533-8 - PubMed

MeSH terms

Substances

LinkOut - more resources