. 2022 Jul;260(7):2271-2281.
doi: 10.1007/s00417-022-05579-9. Epub 2022 Feb 16.

Repeatability of a new swept-source optical coherence tomographer and agreement with other three optical biometers

Affiliations

Repeatability of a new swept-source optical coherence tomographer and agreement with other three optical biometers

Shi-Ming Cheng et al. Graefes Arch Clin Exp Ophthalmol. 2022 Jul.

Abstract

Purpose: To investigate the repeatability of Anterion and compare the agreement of ocular biometric measurements and predicted intraocular lens (IOL) powers with other three optical biometers.

Methods: Flat keratometry (Kf), steep keratometry (Ks), J0 and 45 vectors, central cornea thickness (CCT), anterior chamber depth (ACD), and axial length (AL) from the Anterion, IOLMaster 700, Lenstar LS 900, and OA-2000 were recorded. The IOL powers were calculated with the Hoffer Q, Holladay 1, SRK/T, and Haigis formulas. The repeatability was evaluated using the within-subject standard deviation (Sw), repeatability coefficient (RC), coefficient of variation (COV), and intraclass correlation coefficient (ICC). Inter-device agreement between the four biometers was assessed with the 95% limits of agreement.

Results: In total, 101 right eyes of 101 participants were enrolled. The Anterion showed good repeatability for all the included biometric parameters with all the CoV ≤ 0.30% and ICC ≥ 0.930 except for J45 with moderate repeatability (ICC was 0.849). Good agreement was found among the four devices for Kf, Ks, J0, J45, ACD, and AL. Generally, wide 95% LoA was found for the predicted IOL powers with the four IOL calculation formulas between the four devices.

Conclusions: The Anterion showed good repeatability of biometric measurements for most parameters. Good agreement among the four optical biometers was achieved for all the parameters except for CCT and the predicted IOL power. The AL values exhibited the best repeatability with Anterion and the best agreement among the biometers in our study.

Keywords: Agreement; Anterion; Anterior segment biometry; Axial length; Optical coherence tomographer; Repeatability.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Tideman J, Polling JR, Vingerling JR, Jaddoe V, Williams C, Guggenheim JA, Klaver C (2018) Axial length growth and the risk of developing myopia in European children. Acta Ophthalmol 96:301–309. https://doi.org/10.1111/aos.13603 - DOI - PubMed
    1. Zhang Y, Chen Y (2019) Effect of orthokeratology on axial length elongation in anisomyopic children. Optom Vis Sci 96:43–47. https://doi.org/10.1097/OPX.0000000000001315 - DOI - PubMed
    1. Day AC, Machin D, Aung T, Gazzard G, Husain R, Chew PT, Khaw PT, Seah SK, Foster PJ (2011) Central corneal thickness and glaucoma in East Asian people. Invest Ophthalmol Vis Sci 52:8407–8412. https://doi.org/10.1167/iovs.11-7927 - DOI - PubMed
    1. Dutta D, Rao HL, Addepalli UK, Vaddavalli PK (2013) Corneal thickness in keratoconus: comparing optical, ultrasound, and optical coherence tomography pachymetry. Ophthalmology 120:457–463. https://doi.org/10.1016/j.ophtha.2012.08.036 - DOI - PubMed
    1. Kumar N, Pop-Busui R, Musch DC, Reed DM, Momont AC, Hussain M, Raval N, Moroi SE, Shtein R (2018) Central corneal thickness increase due to stromal thickening with diabetic peripheral neuropathy severity. Cornea 37:1138–1142. https://doi.org/10.1097/ICO.0000000000001668 - DOI - PubMed - PMC

LinkOut - more resources