Correlation between bispectral index and patient state index in children under sevoflurane anesthesia
- PMID: 35191119
- DOI: 10.1111/pan.14422
Correlation between bispectral index and patient state index in children under sevoflurane anesthesia
Abstract
Background: Because the unanticipated arousal or hemodynamic instability during anesthesia may adversely affect the physical and emotional welfare of children, adequate management of the anesthesia depth is required. We aimed to compare Bispectral Index (BIS) and Patient State Index (PSI) in children during sevoflurane anesthesia and evaluate PSI as depth of anesthesia monitor in children aged 6 months-12 years.
Methods: In this prospective observational study, children aged 6 months-12 years old scheduled for elective surgery under sevoflurane anesthesia were enrolled from November 2018 to June 2019. We monitored BIS and PSI at different sevoflurane concentrations. The primary outcome was the correlation between BIS and PSI. The correlation between BIS and PSI at different sevoflurane concentrations (at 1, 1.5, and 2 MACs) and at different age groups (6 months-2 years, 2-7 years, and 8-12 years) was also investigated.
Results: Bispectral index and PSI showed a fair correlation (r = .430; 95% confidence interval [CI], 0.297-0.546; p < .001). Two values were fairly correlated at 1, 1.5, and 2 MAC (r = .544; 95% CI, 0.314-0.716; p < .001, r = .509; 95% CI, 0.283-0.699; p < .001, and r = .315; 95% CI, 0.047-0.522; p = 0.007). BIS and PSI values showed a fair correlation in 6 months - 2 year and 8-12 year groups (r = .696; 95% CI, 0.519-0.813; p < .001 and r = .297; 95% CI, -0.017 to 0.543; p < .021), but there was not significant correlation in 2-7 years group (r = .190; 95% CI, -0.015 to 0.374; p = .052).
Conclusions: There was a fair correlation between BIS and PSI in children under sevoflurane anesthesia. The use of BIS and PSI as an indicator for anesthesia depth by sevoflurane is not reliable in pediatric patients.
© 2022 John Wiley & Sons Ltd.
References
REFERENCES
-
- Bruhn J, Myles PS, Sneyed R. Depth of anaesthesia monitoring: what’s available, what’s validated and what’s next? Br J Anaesth. 2006;97:85-94.
-
- Rampil IJ. A primer for EEG signal processing in anesthesia. Anesthesiology. 1998;89:980-1002.
-
- Prichep L, Gugino L, John E, et al. The patient state index as an indicator of the level of hypnosis under general anaesthesia. Br J Anaesth. 2004;92:393-399.
-
- Schneider G, Mappes A, Neissendorfer T, Schabacker M, Kuppe H, Kochs E. EEG-based indices of anaesthesia: correlation between bispectral index and patient state index? Eur J Anaesthesiol. 2004;21:6-12.
-
- Soehle M, Ellerkmann RK, Grube M, et al. Comparison between bispectral index and patient state index as measures of the electroencephalographic effects of sevoflurane. Anesthesiology. 2008;109:799-805.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous