Fatigue Resistance of Cast-on Implant Abutment Fabricated with Three Different Alloys
- PMID: 35196721
- PMCID: PMC9949937
- DOI: 10.1055/s-0041-1742124
Fatigue Resistance of Cast-on Implant Abutment Fabricated with Three Different Alloys
Abstract
Objectives: This study aimed to evaluate fatigue resistance of cast-on implant abutment using three alloys.
Materials and methods: Forty specimens of implant-supported crowns were prepared; Group 1 (TA) stock titanium abutments, Group 2 (GS) abutment cast with 40% gold alloy, Group 3 (GP) abutment cast with palladium alloy, and Group 4 (CN) abutment cast with nickel-chromium alloy. Specimens were cyclic loaded at 20 Hz, starting from 200 N (5,000 cycles), followed by stepwise loading of 400, 600, 800, 1,000, 1,200, 1,400, 1,600, and 1,800 N (30,000 cycles/step). Specimens were loaded until failure or reached 245,000 cycles.
Statistical analysis: The withstand cycles were analyzed using one-way analysis of variance and Weibull survival analysis. Fracture surfaces were examined using scanning electron microscopy.
Results: The results of withstand cycles were TA (189,883 ± 22,734), GS (195,028 ± 22,371), GP (187,662 ± 22,555), and CN (200,350 ± 30,851). The statistical analysis showed no significant difference between the groups (p = 0.673).
Conclusion: Although CN has higher Weibull characteristic strength which means greater durability, its lower Weibull modulus demonstrated less structural reliability. Consistent failures at implant fixture level were also found in CN group.
The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. (https://creativecommons.org/licenses/by/4.0/).
Conflict of interest statement
None declared.
Figures





References
-
- Barbosa G AS, das Neves F D, de Mattos MdaGC, Rodrigues R CS, Ribeiro R F. Implant/abutment vertical misfit of one-piece cast frameworks made with different materials. Braz Dent J. 2010;21(06):515–519. - PubMed
-
- Hulterström M, Nilsson U. Cobalt-chromium as a framework material in implant-supported fixed prostheses: a preliminary report. Int J Oral Maxillofac Implants. 1991;6(04):475–480. - PubMed
-
- Barbosa G AS, Simamoto Júnior P C, Fernandes Neto A J, de Mattos MdaGC, das Neves F D. Prosthetic laboratory influence on the vertical misfit at the implant/UCLA abutment interface. Braz Dent J. 2007;18(02):139–143. - PubMed
-
- Woodward B. Palladium in temporary and permanently implantable medical devices. Johnson Matthey. 2012;56(03):213–217.
Grants and funding
LinkOut - more resources
Full Text Sources
Miscellaneous