The digital scent device as a new concept for olfactory assessment

Affiliations

The digital scent device as a new concept for olfactory assessment

Marcio Nakanishi et al. Int Forum Allergy Rhinol. 2022 Oct.

Abstract

Background: There are major challenges in olfactory measurements in clinical practice; therefore, a handheld digital scent device (DSD; Noar MultiScent 20) was developed as a tablet with an integrated storage system for odors. The DSD is a self-administered, handheld device that controls the duration of odor release to the nasal cavity through a touchscreen digital interface with automatic database generation. In this study we aimed to determine the feasibility of this DSD as an olfactory assessment test.

Methods: We recruited 180 participants (age [mean ± standard deviation], 34.58 ± 9.71 years; 114 women and 66 men) to participate in smell tests using both the DSD and the 40-item Smell Identification Test (SIT-40), which contained the same type and order of odors and the same multiple-choice answers. The scores were compared and evaluated for correlation between the tests, and test-retest reliability was calculated.

Results: The DSD test scores were higher than the SIT-40 scores (median [interquartile range], 32 [5.0] vs 31 [7.0]; p = 0.005). The completion time was less for the DSD test than for the SIT-40 (12.5 [5.0] vs 16 [6.0] minutes; p < 0.001). The tests were strongly correlated (Spearman rho = 0.74; p < 0.001) and exhibited a high level of agreement (Bland-Altman regression coefficient = 0.672; p = 0.003). The DSD test-retest reliability coefficient was 0.820.

Conclusion: The DSD is feasible as an olfactory assessment test. The digitalization of olfactory assessment combined with data science may enable new research perspectives in the field of olfaction.

Keywords: anosmia; hyposmia; odor; olfaction; reliability; scent; smell.

PubMed Disclaimer

References

REFERENCES

    1. Doty RL, Shaman P, Dann M. Development of the university of Pennsylvania Smell Identification Test: a standardized microencapsulated test of olfactory function. Physiol Behav. 1984;32:489-502.
    1. Cain WS, Gent J, Catalanotto FA, Goodspeed RB. Clinical evaluation of olfaction. Am J Otolaryngol. 1983;4:252-256.
    1. Hummel T, Sekinger B, Wolf SR, et al. ‘Sniffin’ sticks': olfactory performance assessed by the combined testing of odor identification, odor discrimination and olfactory threshold. Chem Senses. 1997;22:39-52.
    1. Su B, Bleier B, Wei Y, Wu D. Clinical implications of psychophysical olfactory testing: assessment, diagnosis, and treatment outcome. Front Neurosci. 2021;15:646956.
    1. Hummel T, Whitcroft KL, Andrews P, et al. Position paper on olfactory dysfunction. Rhinology. 2017;54(suppl):1-30.

Publication types

LinkOut - more resources