A Ten-Year Radiostereometric Analysis of Polyethylene Wear Between Oxidized Zirconium and Cobalt Chrome Articulations in Total Hip Arthroplasty
- PMID: 35245689
- DOI: 10.1016/j.arth.2022.02.099
A Ten-Year Radiostereometric Analysis of Polyethylene Wear Between Oxidized Zirconium and Cobalt Chrome Articulations in Total Hip Arthroplasty
Abstract
Background: Oxidized zirconium (OxZi) femoral heads combine the decreased abrasive properties of ceramics with the toughness of metal alloys to lower wear and increase the durability of total hip arthroplasty. The purpose of this study was to compare the 10-year polyethylene wear rates between OxZi/XLPE and cobalt chrome (CoCr)/XLPE articulations.
Methods: A radiostereometric analysis (RSA) was performed on two cohorts of 23 patients who underwent total hip arthroplasty using either OxZi/XLPE or CoCr/XLPE at a minimum of 10-year follow-up. Cohorts were matched for age, gender, body mass index (BMI), and diagnosis. Polyethylene wear was measured using RSA to determine total and steady-state wear rates for both cohorts. Preoperative and postoperative patient-reported outcome measures (SF12, HHS, and Western Ontario and McMaster Universities Arthritis Index scores) were compared.
Results: The mean total head penetration rate was found to be statistically different between the entire cohorts (OxZi 0.048 ± 0.021 mm/y, CoCr 0.035 ± 0.017 mm/y, P = .02) but not when 28-mm heads only (OxZi 0.045 ± 0.016 mm/y, CoCr 0.034 ± 0.017 mm/y, P = .066) were directly compared. The mean steady-state wear rate was not significantly different between the entire cohorts (OxZi 0.031 ± 0.021 mm/y, CoCr 0.024 ± 0.019 mm/y, P = .24) or 28-mm head cohorts (OxZi 0.028 ± 0.019 mm/y, CoCr 0.024 ± 0.019 mm/y, P = .574). Outcome measures showed no statistical difference except for the Harris Hip Score where the OxZi cohort demonstrated higher median scores.
Conclusion: Using RSA to evaluate the 10-year in-vivo head penetration, there was no statistically significant difference in steady-state wear rates between OxZi and CoCr articulations. Both bearing combinations demonstrated wear rates well below the threshold for osteolysis.
Keywords: cobalt chromium; femoral head; highly-crosslinked polyethylene; oxidized zirconium; total hip arthroplasty; wear.
Copyright © 2022 Elsevier Inc. All rights reserved.
Similar articles
-
Five-year comparison of wear using oxidised zirconium and cobalt-chrome femoral heads in total hip arthroplasty: a multicentre randomised controlled trial.Bone Joint J. 2015 Jul;97-B(7):883-9. doi: 10.1302/0301-620X.97B7.35285. Bone Joint J. 2015. PMID: 26130341 Clinical Trial.
-
Long-term polyethylene wear rates and clinical outcomes of oxidized zirconium femoral heads on highly cross-linked polyethylene in total hip arthroplasty.Bone Joint J. 2024 Mar 1;106-B(3 Supple A):38-43. doi: 10.1302/0301-620X.106B3.BJJ-2023-0836.R1. Bone Joint J. 2024. PMID: 38423114
-
Oxidized zirconium versus cobalt-chrome femoral heads in total hip arthroplasty: a multicentre prospective randomized controlled trial with ten years' follow-up.Bone Joint J. 2022 Jul;104-B(7):833-843. doi: 10.1302/0301-620X.104B7.BJJ-2021-1673.R1. Bone Joint J. 2022. PMID: 35775177 Clinical Trial.
-
Is Oxidized Zirconium Femoral Head Superior to Other Bearing Types in Total Hip Arthroplasty? A Systematic Review and Meta-Analysis.J Arthroplasty. 2019 Aug;34(8):1844-1852. doi: 10.1016/j.arth.2019.03.072. Epub 2019 Apr 2. J Arthroplasty. 2019. PMID: 31010775
-
Ceramic Femoral Heads Exhibit Lower Wear Rates Compared to Cobalt Chrome: A Meta-Analysis.J Arthroplasty. 2023 Feb;38(2):397-405. doi: 10.1016/j.arth.2022.09.008. Epub 2022 Sep 12. J Arthroplasty. 2023. PMID: 36108994
Cited by
-
Ultra-High-Molecular-Weight Polyethylene in Hip and Knee Arthroplasties.Materials (Basel). 2023 Mar 7;16(6):2140. doi: 10.3390/ma16062140. Materials (Basel). 2023. PMID: 36984020 Free PMC article. Review.
-
[Physiological reactions in the interface between cementless implants and bone].Orthopadie (Heidelb). 2023 Mar;52(3):178-185. doi: 10.1007/s00132-023-04347-9. Epub 2023 Feb 7. Orthopadie (Heidelb). 2023. PMID: 36749380 Review. German.
-
Polyethylene wear and cup migration of cemented total hip arthroplasty with femoral heads made of oxidized zirconium, steel, or cobalt chromium: a 10-year secondary analysis from a randomized trial using radiostereometry.Acta Orthop. 2024 Sep 30;95:578-585. doi: 10.2340/17453674.2024.41945. Acta Orthop. 2024. PMID: 39347798 Free PMC article. Clinical Trial.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical