. 2022 May;33(5):874-882.
doi: 10.1111/jce.15442. Epub 2022 Mar 21.

In vivo Lesion Index (LSI) validation in percutaneous radiofrequency catheter ablation

Affiliations
Free article

In vivo Lesion Index (LSI) validation in percutaneous radiofrequency catheter ablation

Sakis Themistoclakis et al. J Cardiovasc Electrophysiol. 2022 May.
Free article

Abstract

Introduction: Lesion Index (LSI) has been developed to predict lesion efficacy during radiofrequency (RF) catheter ablation. However, its value in predicting lesions size has still to be established. The aim of our study was to assess the lesions size reproducibility for prespecified values of LSI reached during RF delivery in an in vivo beating heart.

Methods: Ablation lesions were created with different values of LSI in seven domestic pigs by means of a contact force-sensing catheter (TactiCathTM , Abbott). Lesions were identified during RF delivery by means of a three-dimensional mapping system (EnSiteTM Precision, Abbott) and measured after heart explantation. Histology was carried out after gross examination on the first three lesions to confirm the accuracy of the macroscopic evaluation.

Results: A total of 64 myocardial lesions were created. Thirty-nine lesions were excluded from the analysis for the following reasons: histological confirmation of macroscopic lesion measurement (n = 3), transmurality (n = 24), unfavorable anatomic position (n = 10), not macroscopically identifiable (n = 2). In a final set of 25 nontransmural lesions, injury width and depth were, respectively, 4.6 ± 0.6 and 2.6 ± 0.8 mm for LSI = 4, 7.3 ± 0.8 and 4.7 ± 0.6 mm for LSI = 5, and 8.6 ± 1.2 and 7.2 ± 1.1 mm for LSI = 6. A strong linear correlation was observed between LSI and lesion width (r = .87, p < .00001) and depth (r = .89, p < .00001). Multiple linear regression analysis identified LSI as the only ablation parameter that significantly predicted lesion width (p < .001) and depth (p < .001).

Conclusion: In our in vivo study, LSI proved highly predictive of lesion size and depth.

Keywords: Lesion Index; arrhythmias; contact force; in vivo model; myocardial lesions; radiofrequency energy.

PubMed Disclaimer

Comment in

References

REFERENCES

    1. Calkins H, Hindricks G, Cappato R, et al. 2017 HRS/EHRA/ECAS/APHRS/SOLAECE expert consensus statement on catheter and surgical ablation of atrial fibrillation. Heart Rhythm. 2017;14:e275-e444.
    1. Raviele A, Natale A, Calkins H, et al. Venice Chart International Consensus Document on Atrial Fibrillation Ablation: 2011 Update. J Cardiovasc Electrophysiol. 2012;23:890-923.
    1. Wittkampf FHM, Nakagawa H. RF catheter ablation: lessons on lesions. Pacing Clin Electrophysiol. 2006;29:1285-1297.
    1. Neuzil P, Reddy VY, Kautzner J, et al. Electrical reconnection after pulmonary vein isolation is contingent on contact force during initial treatment. Results from the EFFICAS I Study. Circ Arrhythm Electrophysiol. 2013;6:327-333.
    1. Kautzner J, Neuzil P, Lambert H, et al. EFFICAS II: optimization of catheter contact force improves outcome of pulmonary vein isolation for paroxysmal atrial fibrillation. Europace. 2015;17:1229-1235.

Publication types

Grants and funding