. 2022 May;128(5):e314-e317.
doi: 10.1016/j.bja.2022.02.007. Epub 2022 Feb 21.

Assessment of a splitter for protective dual-patient ventilation in patients with acute respiratory distress syndrome

Affiliations

Assessment of a splitter for protective dual-patient ventilation in patients with acute respiratory distress syndrome

Daniel H Arellano et al. Br J Anaesth. 2022 May.
No abstract available

Keywords: COVID-19; dual-patient ventilation; mechanical ventilation; protective ventilation; respiratory mechanics.

PubMed Disclaimer

Figures

Fig 1
Fig 1
Tidal volume, plateau and peak pressures, and PEEP during dual-patient ventilation without or with a splitter. Median and inter-quartile range of PEEP, plateau pressure (PPL), peak pressure (Ppeak), and tidal volume showing changes during dual-patient ventilation (DPV) under different combinations of airway resistance (Raw) and compliance (Crs) (R5/C75, R5/C60, R5/C25, R20/C75, R20/C60, and R20/C25), and during air leak or airway obstruction. (a) DPV with or without a splitter was delivered to two test lungs (lung A (dashed line) and B (solid line). (b) DPV with a splitter was delivered to a patient (dashed line) and a test lung (solid line). (c) DPV with a splitter was delivered to two patients (dashed and solid lines) for 3 h. Friedman test P-values are displayed for the lung A or patient.
Fig 1
Fig 1
Tidal volume, plateau and peak pressures, and PEEP during dual-patient ventilation without or with a splitter. Median and inter-quartile range of PEEP, plateau pressure (PPL), peak pressure (Ppeak), and tidal volume showing changes during dual-patient ventilation (DPV) under different combinations of airway resistance (Raw) and compliance (Crs) (R5/C75, R5/C60, R5/C25, R20/C75, R20/C60, and R20/C25), and during air leak or airway obstruction. (a) DPV with or without a splitter was delivered to two test lungs (lung A (dashed line) and B (solid line). (b) DPV with a splitter was delivered to a patient (dashed line) and a test lung (solid line). (c) DPV with a splitter was delivered to two patients (dashed and solid lines) for 3 h. Friedman test P-values are displayed for the lung A or patient.
Fig. S1
Fig. S1

References

    1. Sommer D.D., Fisher J.A., Ramcharan V., Marshall S., Vidic D.M. Improvised automatic lung ventilation for unanticipated emergencies. Crit Care Med. 1994;22:705–709. - PubMed
    1. Neyman G., Irvin C.B. A single ventilator for multiple simulated patients to meet disaster surge. Acad Emerg Med. 2006;13:1246–1249. - PMC - PubMed
    1. Chatburn R.L., Branson R.D., Hatipoğlu U. Multiplex ventilation: a simulation-based study of ventilating 2 patients with a single ventilator. Respir Care. 2020;65:920–931. - PubMed
    1. Beitler J.R., Mittel A.M., Kallet R., et al. Ventilator sharing during an acute shortage caused by the COVID-19 pandemic. Am J Respir Crit Care Med. 2020;202:600–604. - PMC - PubMed
    1. Laffey J.G., Chikhani M., Bates D.G., Hardman J.G. Supporting more than one patient with a single mechanical ventilator: useful last resort or unjustifiable risk? Br J Anaesth. 2020;125:247–250. - PMC - PubMed