Prospective validation of gas man simulations of sevoflurane in O2/air over a wide fresh gas flow range
Abstract
The use of inhaled anesthetics has come under increased scrutiny because of their environmental effects. This has led to a shift where sevoflurane in O2/air has become the predominant gas mixture to maintain anesthesia. To further reduce environmental impact, lower fresh gas flows (FGF) should be used. An accurate model of sevoflurane consumption allows us to assess and quantify the impact of the effects of lowering FGFs. This study therefore tested the accuracy of the Gas Man® model by determining its ability to predict end-expired sevoflurane concentrations (FETsevo) in patients using a protocol spanning a wide range of FGF and vaporizer settings. After IRB approval, 28 ASA I-II patients undergoing a gynecologic or urologic procedure under general endotracheal anesthesia were enrolled. Anesthesia was maintained with sevoflurane in O2/air, delivered via a Zeus or FLOW-i workstation (14 patients each). Every fifteen min, FGF was changed to randomly selected values ranging from 0.2 to 6 L/min while the sevoflurane vaporizer setting was left at the discretion of the anesthesiologist. The FETsevo was collected every min for 1 h. For each patient, a Gas Man® simulation was run using patient weight and the same FGF, vaporizer and minute ventilation settings used during the procedure. For cardiac output, the Gas Man default setting was used (= Brody formula). Gas Man®'s performance was assessed by comparing measured with Gas Man® predicted FETsevo using linear regression and Varvel's criteria [median performance error (MDPE), median absolute performance error (MDAPE), and divergence]. Additional analysis included separating performance for the wash-in (0-15 min) and maintenance phase (15-60 min). For the FLOW-i, MDPE, MDAPE and divergence were 1% [- 6, 8], 7% [3, 15] and - 0.96%/h [- 1.14, - 0.88], respectively. During the first 15 min, MDPE and MDAPE were 18% [1, 51] and 21% [8, 51], respectively, and during the last 45 min 0% [- 7, 5] and 6% [2, 10], respectively. For the Zeus, MDPE, MDAPE and divergence were 0% [- 5, 8], 6% [3, 12] and - 0.57%/h [- 0.85, - 0.16], respectively. During the first 15 min, MDPE and MDAPE were 7% [- 6, 28] and 13% [6, 32], respectively, and during the last 45 min - 1% [- 5, 5] and 5% [2, 9], respectively. In conclusion, Gas Man® predicts FETsevo in O2/air in adults over a wide range of FGF and vaporizer settings using different workstations with both MDPE and MDAPE < 10% during the first hour of anesthesia, with better relative performance for simulating maintenance than wash-in. In the authors' opinion, this degree of performance suffices for Gas Man® to be used to quantify the environmental impact of FGF reduction in real life practice of the wash-in and maintenance period combined.
Keywords: Gas Man; Inhaled anesthetics; Low flow anesthesia; Sevoflurane.
© 2022. The Author(s), under exclusive licence to Springer Nature B.V.
Similar articles
-
Development and validation of a model to calculate anesthetic agent consumption from inspired and end-expired concentrations, minute ventilation, fresh gas flow and dead space ventilation.J Clin Monit Comput. 2023 Feb;37(1):227-235. doi: 10.1007/s10877-022-00883-5. Epub 2022 Jun 16. J Clin Monit Comput. 2023. PMID: 35708832
-
Memsorb™, a novel CO2 removal device part I: in vitro performance with the Zeus IE®.J Clin Monit Comput. 2022 Dec;36(6):1591-1600. doi: 10.1007/s10877-021-00802-0. Epub 2022 Jan 28. J Clin Monit Comput. 2022. PMID: 35089526
-
The effect of different flow levels and concentrations of sevoflurane during the wash-in phase on volatile agent consumption: a randomized controlled trial.J Clin Monit Comput. 2022 Oct;36(5):1257-1262. doi: 10.1007/s10877-022-00846-w. Epub 2022 Apr 19. J Clin Monit Comput. 2022. PMID: 35438367 Clinical Trial.
-
Special aspects of pharmacokinetics of inhalation anesthesia.Handb Exp Pharmacol. 2008;(182):159-86. doi: 10.1007/978-3-540-74806-9_8. Handb Exp Pharmacol. 2008. PMID: 18175091 Review.
-
[Low-flow and minimal-flow anesthesia with sevoflurane].Anaesthesist. 1998 Nov;47 Suppl 1:S70-6. doi: 10.1007/pl00002503. Anaesthesist. 1998. PMID: 9893885 Review. German.
Cited by
-
Preventing Prolonged Times to Awakening While Mitigating the Risk of Patient Awareness: Gas Man Computer Simulations of Sevoflurane Consumption From Brief, High Fresh Gas Flow Before the End of Surgery.Cureus. 2024 Mar 6;16(3):e55626. doi: 10.7759/cureus.55626. eCollection 2024 Mar. Cureus. 2024. PMID: 38586680 Free PMC article.
-
Enhancing Anesthesia Education and Clinical Practice: A Comprehensive Review of GASMAN Simulation Software.J Med Educ Curric Dev. 2024 Sep 17;11:23821205241283804. doi: 10.1177/23821205241283804. eCollection 2024 Jan-Dec. J Med Educ Curric Dev. 2024. PMID: 39314826 Free PMC article. Review.
References
-
- Sulbaek Andersen MP, Nielsen OJ, Wallington TJ, Karpichev B, Sander SP. Medical intelligence article: assessing the impact on global climate from general anesthetic gases. Anesth Analg. 2012;114:1081–5. https://doi.org/10.1213/ANE.0b013e31824d6150 . - DOI - PubMed
-
- Hönemann C, Kim SC. Please stop using nitrous oxide in routine clinical practice (comment on: use of nitrous oxide in contemporary anesthesia-an ongoing tug of war). Can J Anaesth. 2021. https://doi.org/10.1007/s12630-021-02136-6 . - DOI - PubMed - PMC
-
- Jahn UR, Berendes E. Nitrous oxide–an outdated anaesthetic. Best Pract Res Clin Anaesthesiol. 2005;19:391–7. https://doi.org/10.1016/j.bpa.2005.03.003 . - DOI - PubMed
-
- Andersen MPS, Nielsen OJ, Sherman JD. The global warming potentials for anesthetic gas sevoflurane need significant corrections. Environ Sci Technol. 2021;55:10189–91. https://doi.org/10.1021/acs.est.1c02573 . - DOI - PubMed
-
- Sherman J, Le C, Lamers V, Eckelman M. Life cycle greenhouse gas emissions of anesthetic drugs. Anesth Analg. 2012;114:1086–90. https://doi.org/10.1213/ANE.0b013e31824f6940 . - DOI - PubMed
MeSH terms
Substances
LinkOut - more resources
Full Text Sources