Gated Tomographic Radionuclide Angiography Using 3D-Ring CZT StarGuide SPECT/CT Head-To-Head Comparison With a Cardiac-Dedicated CZT Camera: First Clinical Use and Validation
- PMID: 35353756
- DOI: 10.1097/RLU.0000000000004153
Gated Tomographic Radionuclide Angiography Using 3D-Ring CZT StarGuide SPECT/CT Head-To-Head Comparison With a Cardiac-Dedicated CZT Camera: First Clinical Use and Validation
Abstract
Gated tomographic radionuclide angiography can assess and monitor left cardiac function. Dedicated cardiac CZT cameras have enabled dose reduction and quicker acquisitions. New 3D-ring CZT general purpose systems are now available. We report 50 patients who underwent a 7-minute acquisition on a cardiac-dedicated CZT camera and 9 minutes on a new 3D-ring CZT system after mean injection of 321.4 ± 55.9 MBq 99mTc-labelled human serum albumin. There was no significant difference in left ventricular volumes, and left and right ventricular ejection fractions. These preliminary results seem to validate the use of 3D-ring CZT system for LEVF and cardiac function evaluation.
Copyright © 2022 Wolters Kluwer Health, Inc. All rights reserved.
Conflict of interest statement
Conflicts of interest and sources of funding: CHR Orleans received financial and material support from GE Healthcare regarding the StarGuide evaluation.
References
-
- de Geus-Oei L-F, Mavinkurve-Groothuis AMC, Bellersen L, et al. Scintigraphic techniques for early detection of cancer treatment-induced cardiotoxicity. J Nucl Med Technol . 2013;41:170–181.
-
- Yeh ETH, Bickford CL. Cardiovascular complications of cancer therapy: incidence, pathogenesis, diagnosis, and management. J Am Coll Cardiol . 2009;53:2231–2247.
-
- Carver JR, Shapiro CL, Ng A, et al. American Society of Clinical Oncology clinical evidence review on the ongoing care of adult cancer survivors: cardiac and pulmonary late effects. J Clin Oncol . 2007;25:3991–4008.
-
- Curigliano G, Cardinale D, Suter T, et al. Cardiovascular toxicity induced by chemotherapy, targeted agents and radiotherapy: ESMO Clinical Practice Guidelines. Ann Oncol . 2012;23(Suppl 7):vii155–vii166.
-
- Einstein AJ, Blankstein R, Andrews H, et al. Comparison of image quality, myocardial perfusion, and left ventricular function between standard imaging and single-injection ultra-low-dose imaging using a high-efficiency SPECT camera: the millisievert study. J Nucl Med . 2014;55:1430–1437.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources