Polysorbate 80 and histidine quantitative analysis by NMR in the presence of virus-like particles

Affiliations

Polysorbate 80 and histidine quantitative analysis by NMR in the presence of virus-like particles

Richard R Rustandi. Electrophoresis. 2022 Jul.

Abstract

Polysorbate-80 (PS80) and histidine are common excipients in vaccine and therapeutic protein formulation. A simple quantitative NMR method to measure both PS80 and histidine in human papillomavirus (HPV) virus-like particle (VLP) vaccine for aqueous and alum-containing samples is described. The new NMR method is compared to current colorimetric methods for PS80 and RP HPLC for histidine. The new NMR method is comparable to current assays with an advantage of a simpler sample treatment for PS80. The efficiency is also increased because one method can now provide two assay results instead of two separate methods. Furthermore, the NMR method can detect PS80 stability due to hydrolysis and oxidation when PS80 is stored in a stainless steel container by observing a change of its NMR line shape profile.

Keywords: HPV; NMR; histidine; polysorbate-80; stainless steel; vaccine.

PubMed Disclaimer

Conflict of interest statement

The author has declared no conflict of interest.

Figures

FIGURE 1
FIGURE 1
Polysorbate‐80 structure
FIGURE 2
FIGURE 2
(A) NMR spectrum of PS80 and histidine in the presence of HPV VLP samples illustrating that no HPV VLP is observed; (B) the enlarged of PS80 and histidine peaks at 3.70 and 7.21 ppm, respectively used for quantitation. (C) Standard curve for histidine. (D) Standard curve for PS80
FIGURE 3
FIGURE 3
Comparison of NMR peak at 3.70 ppm of PS80 (0.015%) in histidine buffer stored in glass (blue trace) and stainless steel (red trace) containers. The PS80 containing HPV VLP sample stored in stainless steel (green trace) and PS80 reference standard control at 0.02% (black trace)
FIGURE 4
FIGURE 4
NMR peak at 3.70 ppm of PS80‐treated with NaOD (red trace) and compared with PS80‐containing HPV VLP stored in stainless steel (green trace) and PS80 reference standard as control (black trace)

References

    1. Zhai L, Tumban E. Gardasil‐9: a global survey of projected efficacy. Antiviral Res. 2016;130:101–9. - PubMed
    1. Zhao Q, Potter CS, Carragher B, Lander G, Sworen J, Towne V, et al. Characterization of virus‐like particles in GARDASIL® by cryo transmission electron microscopy. Hum Vaccin Immunother. 2014;10:734–9. - PMC - PubMed
    1. Gopalrathnam G, Sharma AN, Dodd SW, Huang L. Impact of stainless steel exposure on the oxidation of polysorbate 80 in histidine placebo and active monoclonal antibody formulation. PDA J Pharm Sci Technol. 2018;72:163–75. - PubMed
    1. Martos A, Koch W, Jiskoot W, Wuchner K, Winter G, Friess W, et al. Trends on analytical characterization of polysorbates and their degradation products in biopharmaceutical formulations. J Pharm Sci. 2017;106:1722–35. - PubMed
    1. Abrar S, Trathnigg B. Analysis of polyethyleneoxide macromonomers by liquid chromatography along the critical adsorption line. Anal Bioanal Chem. 2011;400:2577–86. - PubMed

LinkOut - more resources