. 2024 Jan;131(1):146-154.
doi: 10.1016/j.prosdent.2022.02.016. Epub 2022 Mar 30.

Effect of scan-path length on the scanning accuracy of completely dentate and partially edentulous maxillae

Affiliations

Effect of scan-path length on the scanning accuracy of completely dentate and partially edentulous maxillae

Moritz Waldecker et al. J Prosthet Dent. 2024 Jan.

Abstract

Statement of problem: The accuracy of fit of fixed partial dentures is directly dependent on the accuracy of a digital scan. However, the influence of scan-path length on scanning accuracy is unclear.

Purpose: The purpose of this in vitro study was to evaluate how scan-path length influenced the scanning accuracy of a completely dentate or partially edentulous maxilla captured by 3 intraoral scanners: Omnicam AC (OC), TRIOS 4 (TR), and Primescan (PS).

Material and methods: Each intraoral scanner was used to make 30 scans each of the 2 clinical scenarios (completely dentate and partially edentulous) simulated with a reference model. The partially edentulous model simulated a maxilla with 6 prepared teeth to support a complete arch fixed partial denture. The missing teeth were then added to create a completely dentate model. The prepared teeth were later used to determine distance, angular, and tooth-axis deviations between the reference model (digitized with high precision before the tests) and the intraoral scans. Data were statistically analyzed by using a linear model or, if not applicable, a type II ANOVA (α=.05).

Results: Distance deviations increased linearly as the scan-path length increased. In contrast, angular and tooth-axis deviations did not increase linearly. All types of deviation differed depending on the scanning system used. Regarding the 90% quantile values, total distance deviations related to scan-path length amounted to 1.31 μm/mm (OC), 1.00 μm/mm (PS), and 1.45 μm/mm (TR) for the completely dentate maxilla and 1.10 μm/mm (OC), 1.46 μm/mm (PS), and 1.40 μm/mm (TR) for the partially edentulous maxilla.

Conclusions: Distance deviations became larger as the scan-path length increased.