Pulse transit time-estimated blood pressure: a comparison of beat-to-beat and intermittent measurement

Affiliations

Pulse transit time-estimated blood pressure: a comparison of beat-to-beat and intermittent measurement

Satoshi Hoshide et al. Hypertens Res. 2022 Jun.

Abstract

Pulse transit time (PTT), which refers to the travel time between two arterial sites within the same cardiac cycle, has been developed as a novel cuffless form of continuous blood pressure (BP) monitoring. The aim of this study was to investigate differences in BP parameters, including BP variability, between those assessed by beat-to-beat PTT-estimated BP (eBPBTB) and those assessed by intermittent PTT-estimated BP at fixed time intervals (eBPINT) in patients suspected of having sleep disordered breathing (SDB). In 330 patients with SDB (average age, 66.8 ± 11.9 years; 3% oxygen desaturation index [ODI], 21.0 ± 15.0/h) from 8 institutes, PTT-estimated BP was continuously recorded during the nighttime. The average systolic eBPBTB, maximum systolic and diastolic eBPBTB, standard deviation (SD) of systolic and diastolic eBPBTB, and coefficient variation (CV) of systolic and diastolic eBPBTB were higher than the respective values of eBPINT (all P < 0.05). Bland-Altman analysis showed a close agreement between eBPBTB and eBPINT in average systolic BP and SD and CV of systolic BP, while there were disagreements in both minimum and maximum values of eBPBTB and eBPINT in patients with high systolic BP (P < 0.05). Although systolic BP variability incrementally increased according to the tertiles of 3%ODI in both eBPBTB and eBPINT (all P < 0.05), there was no difference in this tendency between eBPBTB and eBPINT. In patients with suspected SDB, the difference between eBPBTB and eBPINT was minimal, and there were disagreements regarding both the minimum and maximum BP. However, there were agreements in regard to the index of BP variability between eBPBTB and eBPINT.

Keywords: Beat-to-beat; Blood pressure variability; Continuous blood pressure monitoring; Pulse transit time.

PubMed Disclaimer

Conflict of interest statement

TK is affiliated with a department endowed by Philips Respironics, ResMed, and Fukuda Denshi.

Figures

Fig. 1
Fig. 1
Bland–Altman plots comparing systolic blood pressure parameters between eBPBTB with eBPINT. eBPBTB indicates beat-to-beat PTT-estimated BP, eBPINT intermittent PTT-estimated BP at fixed time intervals

References

    1. Scheer B, Perel A, Pfeiffer UJ. Clinical review: complications and risk factors of peripheral arterial catheters used for haemodynamic monitoring in anaesthesia and intensive care medicine. Crit Care. 2002;6:199–204. doi: 10.1186/cc1489. - DOI - PMC - PubMed
    1. Bogert LW, van Lieshout JJ. Non-invasive pulsatile arterial pressure and stroke volume changes from the human finger. Exp Physiol. 2005;90:437–46. doi: 10.1113/expphysiol.2005.030262. - DOI - PubMed
    1. Kario K, Chirinos JA, Townsend RR, Weber MA, Scuteri A, Avolio A, et al. Systemic hemodynamic atherothrombotic syndrome (SHATS)—Coupling vascular disease and blood pressure variability: Proposed concept from pulse of Asia. Prog Cardiovasc Dis. 2020;63:22–32. doi: 10.1016/j.pcad.2019.11.002. - DOI - PubMed
    1. Kario K. Evidence and perspectives on the 24-h management of hypertension: hemodynamic biomarker-initiated ‘anticipation medicine’ for zero cardiovascular event. Prog Cardiovasc Dis. 2016;59:262–81. doi: 10.1016/j.pcad.2016.04.001. - DOI - PubMed
    1. Kario K, Hoshide S, Mizuno H, Kabutoya T, Nishizawa M, Yoshida T, et al. Nighttime blood pressure phenotype and cardiovascular prognosis: practitioner-based nationwide JAMP study. Circulation. 2020;142:1810–20. doi: 10.1161/CIRCULATIONAHA.120.049730. - DOI - PMC - PubMed

LinkOut - more resources