Volumetric wireless coil for wrist MRI at 1.5 T as a practical alternative to Tx/Rx extremity coil: a comparative study
- PMID: 35397309
- DOI: 10.1016/j.jmr.2022.107209
Volumetric wireless coil for wrist MRI at 1.5 T as a practical alternative to Tx/Rx extremity coil: a comparative study
Abstract
This work performs a detailed assessment of radiofrequency (RF) safety and imaging performance of a volumetric wireless coil based on periodically coupled split-loop resonators (SLRs) for 1.5 T wrist MRI versus a commercially available transceive extremity coil. In particular, we evaluated the transmit efficiency and RF safety for three setups: a whole-body birdcage coil, a transceive extremity birdcage coil, and a volumetric wireless coil inductively coupled to the whole-body birdcage coil. The imaging performance of the two latter setups was studied experimentally for nine subjects. The signal-to-noise ratio (SNR) of the images acquired with several standard pulse sequences for osteoarthritis wrist imaging was assessed. Application of the wireless coil significantly improved the specific absorption rate (SAR) efficiency of the whole-body birdcage coil, with at least 4.3-fold and 7.6-fold improvement of local and global SAR efficiencies, respectively. This setup also outperformed the transceive extremity coil in terms of SNR (up to 1.40-fold gain) with a moderate (11%) reduction of the local SAR efficiency.
Keywords: Metamaterials; Radiofrequency coils; Radiofrequency safety; SNR enhancement; Specific absorption rate; Transmit efficiency; Wireless coil; Wrist MRI.
Copyright © 2022 Elsevier Inc. All rights reserved.
Conflict of interest statement
Declaration of Competing Interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous