. 2022 Jun;45(2):577-587.
doi: 10.1007/s13246-022-01123-5. Epub 2022 Apr 11.

Clinical implementation of HyperArc

Affiliations

Clinical implementation of HyperArc

Felix H C Wong et al. Phys Eng Sci Med. 2022 Jun.

Abstract

The aim of this study was to present our experience on clinical implementation of HyperArc including dosimetric comparison between VMAT and HyperArc plans and dosimetric verification of HyperArc. In this study, eleven previously treated cases of brain metastasis were selected from our brain stereotactic radiotherapy program. The cases were retrospectively planned using HyperArc technique and the plan quality was evaluated. In addition, dosimetric effects of HyperArc plan with different energies and using jaw tracking technique were evaluated. Furthermore, dosimetric verification of HyperArc plans was performed using ion chamber and radiochromic film. Our results of dosimetric comparison shows that HyperArc technique improved both conformity index and gradient index compared to VMAT plans. We also found that using 6MV flattening filter free (6MV-FFF) beam improves gradient index in HyperArc plans compared to using 6MV flattening filter beam. Furthermore, our results show that jaw tracking technique reduces the size of low dose volume while maintaining similar target coverage, conformity index, and gradient index. In our dosimetric verification study, results of ion chamber and film measurement indicate no significant difference between VMAT and HyperArc plans. In conclusion, HyperArc simplifies planning of stereotactic treatment for brain and improves the dosimetry in treatment plans. Additionally, HyperArc provides for a safe and efficient treatment delivery system for stereotactic treatments to brain.

Keywords: HyperArc; Stereotactic radiosurgery; Stereotactic radiotherapy.

PubMed Disclaimer

References

    1. Gu L et al (2019) Stereotactic radiation therapy (SRT) for brain metastases of multiple primary tumors: a single institution retrospective analysis. Front Oncol 9:1352. https://doi.org/10.3389/fonc.2019.01352 - DOI - PubMed - PMC
    1. Tsao MN et al (2012) Radiotherapeutic and surgical management for newly diagnosed brain metastasis(es): an american society for radiation oncology evidence-based guideline. Pract Radiat Oncol 2(3):210–225. https://doi.org/10.1016/j.prro.2011.12.004 - DOI - PubMed - PMC
    1. Eichler AF, Loeffler JS (2007) Multidisciplinary management of brain metastases. Oncologist 12(7):884–898. https://doi.org/10.1634/theoncologist.12-7-884 - DOI - PubMed
    1. Siddiqui ZA et al (2020) Predictors of radiation necrosis in long-term survivors after Gamma Knife stereotactic radiosurgery for brain metastases. Neurooncol Pract 7(4):400–408. https://doi.org/10.1093/nop/npz067 - DOI - PubMed
    1. Manning MA et al (2000) Hypofractionated stereotactic radiotherapy as an alternative to radiosurgery for the treatment of patients with brain metastases. Int J Radiat Oncol Biol Phys 47(3):603–608. https://doi.org/10.1016/s0360-3016(00)00475-2 - DOI - PubMed

MeSH terms