Insurance lesions: Does a second lesion make a difference?
Abstract
Introduction: In radiofrequency ablation procedures for cardiac arrhythmia, the efficacy of creating repeated lesions at the same location ("insurance lesions") remains poorly studied. We assessed the effect of type of tissue, power, and time on the resulting lesion geometry during such multiple ablation procedures.
Methods: A custom ex vivo ablation model was used to assess lesion formation. An ablation catheter was oriented perpendicular to the tissue and used to create lesions that varied by type of tissue (atrial or ventricular free wall), power (30 or 50 W), and time (30, 40, or 50 s for standard ablations and 5, 10, or 15 s for high-power, short-duration [HPSD] ablations). Lesion dimensions were recorded and then analyzed. Radiofrequency ablations were performed on 57 atrial tissue samples (28 HPSD, 29 standard) and 28 ventricular tissue samples (all standard).
Results: With ablation parameters held constant, performing multiple ablations significantly increased lesion depth in ventricular tissue when ablations were performed at 30 W for 50 s. No other set of ablation parameters was shown to affect the width or depth of the resulting lesions in either tissue type.
Conclusion: Multiple ablations created with the same power and time, delivered within 30 s of each other at the same exact location, offer no meaningful benefit in lesion depth or width over single ablations, with the exception of ventricular ablation at 30 W for 50 s. Given the risks associated with excessive ablation, our results suggest that this practice should be re-evaluated by clinical electrophysiologists.
Keywords: arrhythmia; catheter ablation/method; insurance lesion; multiple ablations.
© 2022 Wiley Periodicals LLC.
Comment in
-
Radiofrequency ablation-to insure or not to insure is the question.J Cardiovasc Electrophysiol. 2022 Jul;33(7):1513-1514. doi: 10.1111/jce.15499. Epub 2022 Apr 27. J Cardiovasc Electrophysiol. 2022. PMID: 35437896 No abstract available.
-
How lesion size grows after applying a second RF pulse on the same site.J Cardiovasc Electrophysiol. 2022 Dec;33(12):2673. doi: 10.1111/jce.15707. Epub 2022 Oct 23. J Cardiovasc Electrophysiol. 2022. PMID: 36251266 No abstract available.
References
REFERENCES
-
- Barkagan M, Leshem E, Shapira-Daniels A, et al. Histopathological characterization of radiofrequency ablation in ventricular scar tissue. JACC Clin Electrophysiol. 2019;5(8):920-931.
-
- Demazumder D, Mirotznik MS, Schwartzman D. Biophysics of radiofrequency ablation using an irrigated electrode. J Interv Card Electrophysiol. 2001;5(4):377-389.
-
- Leshem E, Zilberman I, Tschabrunn CM, et al. High-power and short-duration ablation for pulmonary vein isolation: biophysical characterization. JACC Clin Electrophysiol. 2018;4(4):467-479.
-
- John M, Rook A, Post A, et al. Bipolar ablation's unique paradigm: duration and power as respectively distinct primary determinants of transmurality and steam pop formation. Heart Rhythm O2. 2020;1(4):290-296.
-
- Kar R, Post A, John M, Rook A, Razavi M. An initial ex vivo evaluation of temperature profile and thermal injury formation on the epiesophageal surface during radiofrequency ablation. J Cardiovasc Electrophysiol. 2021;32(3):704-712.