Significance of quantitative enzyme-linked immunosorbent assay (ELISA) results in evaluation of three ELISAs and Western blot tests for detection of antibodies to human immunodeficiency virus in a high-risk population
- PMID: 3546369
- PMCID: PMC265907
- DOI: 10.1128/jcm.25.2.395-400.1987
Significance of quantitative enzyme-linked immunosorbent assay (ELISA) results in evaluation of three ELISAs and Western blot tests for detection of antibodies to human immunodeficiency virus in a high-risk population
Abstract
The characteristics of primary (first) tests with three enzyme-linked immunosorbent assay (ELISA) kits for human immunodeficiency virus (HIV) antibody were determined. The three ELISAs were performed on 3,229, 3,130, and 685 specimens from high-risk individuals using the Litton (LT; Litton Bionetics Laboratory Products, Charleston, S.C.), Dupont (DP; E. I. du Pont de Nemours & Co., Inc., Wilmington, Del.), and Genetic Systems (GS; Genetic Systems, Seattle, Wash.) kits, respectively. Evaluation was based on the distribution of quantitative test results (such as optical densities), a comparison with Western blot (WB) results, reproducibility of the tests, and identification of seroconverters. The performances of the GS and the DP kits were good by all four criteria and exceeded that of the LT kit. Primary ELISA-negative results were not always confirmed with repeat ELISA and by WB testing. The largest percentage of these unconfirmed negative test results came from samples with quantitative results in the fifth percentile nearest the cutoff. Thus, supplementary testing was indicated for samples with test results in this borderline negative range. Similarly, borderline positive primary ELISA results that were quantitatively nearest (fifth percentile) the cutoff value were more likely to be antibody negative on supplementary testing than samples with high antibody values. In this study, results of repeated tests by GS ELISA showed the least change from first test results. DP ELISA showed more unconfirmed primary positive test results, and LT ELISA showed more unconfirmed primary negative test results. Designation of a specimen with a single ELISA quantitative level near the cutoff value as positive or negative should be viewed with skepticism. A higher than normal proportion of specimens with high negative optical densities by GS ELISA (fifth percentile nearest the cutoff) and also negative by WB were found to be from individuals in the process of seroconversion.
Similar articles
-
Variations in Western blot banding patterns of human T-cell lymphotropic virus type III/lymphadenopathy-associated virus.J Clin Microbiol. 1987 Jan;25(1):81-4. doi: 10.1128/jcm.25.1.81-84.1987. J Clin Microbiol. 1987. PMID: 3540002 Free PMC article.
-
Detection of early antibodies in human immunodeficiency virus infection by enzyme-linked immunosorbent assay, Western blot, and radioimmunoprecipitation.J Clin Microbiol. 1987 Sep;25(9):1605-10. doi: 10.1128/jcm.25.9.1605-1610.1987. J Clin Microbiol. 1987. PMID: 3477569 Free PMC article.
-
[Comparison of two ELISA kits for the demonstration of antibodies against LAV/HTLV-III].Zentralbl Bakteriol Mikrobiol Hyg A. 1987 Apr;264(1-2):196-200. Zentralbl Bakteriol Mikrobiol Hyg A. 1987. PMID: 3307232 German.
-
Status of current clinical tests for human immunodeficiency virus (HIV): applications and limitations.Ann Clin Lab Sci. 1987 Sep-Oct;17(5):279-85. Ann Clin Lab Sci. 1987. PMID: 3314657 Review.
-
[HIV (HTLV III/LAV) serology: experiences based on more than 42,000 tests].Wien Klin Wochenschr. 1987 Aug 7;99(15):536-9. Wien Klin Wochenschr. 1987. PMID: 3310415 Review. German.
Cited by
-
Co-delivery of GPI-anchored CCL28 and influenza HA in chimeric virus-like particles induces cross-protective immunity against H3N2 viruses.J Control Release. 2016 Jul 10;233:208-19. doi: 10.1016/j.jconrel.2016.05.021. Epub 2016 May 10. J Control Release. 2016. PMID: 27178810 Free PMC article.
-
Organizing for ontological change: The kernel of an AIDS research infrastructure.Soc Stud Sci. 2015 Apr;45(2):214-41. doi: 10.1177/0306312714558136. Soc Stud Sci. 2015. PMID: 26477206 Free PMC article.
-
Utility of quantitative enzyme immunoassay reactivity for predicting human immunodeficiency virus seropositivity in low- and high-prevalence populations.J Clin Microbiol. 1994 Jan;32(1):220-3. doi: 10.1128/jcm.32.1.220-223.1994. J Clin Microbiol. 1994. PMID: 8126185 Free PMC article.
-
Are Confirmatory Assays Reliable for HIV-1/HIV-2 Infection Differentiation? A Multicenter Study.J Clin Microbiol. 2023 Aug 23;61(8):e0061923. doi: 10.1128/jcm.00619-23. Epub 2023 Jul 17. J Clin Microbiol. 2023. PMID: 37458587 Free PMC article.
-
Oral fluids as an alternative to serum for measurement of markers of immune activation.Clin Diagn Lab Immunol. 1998 Jul;5(4):507-12. doi: 10.1128/CDLI.5.4.507-512.1998. Clin Diagn Lab Immunol. 1998. PMID: 9665958 Free PMC article.
References
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources