Diagnostic Value of Advanced Metal Artifact Reduction Magnetic Resonance Imaging for Periprosthetic Joint Infection
- PMID: 35467584
- DOI: 10.1097/RCT.0000000000001297
Diagnostic Value of Advanced Metal Artifact Reduction Magnetic Resonance Imaging for Periprosthetic Joint Infection
Abstract
Materials and methods: Magnetic resonance imaging around metal joint prostheses including multiacquisition variable-resonance image combination selective at 1.5 T (from April 2014 to August 2020) was retrospectively evaluated by 2 radiologists for detection of abnormal findings (joint effusion, capsular thickening, pericapsular edema, soft-tissue fluid collection, soft-tissue edema, bone marrow edema pattern around the implant [BME pattern], lymphadenopathy, and others) and overall image impression for PJI. Regarding the soft-tissue fluid collection, presence of communication to the joint or capsular-like structure was evaluated. Clinical assessments were recorded. Positive predictive values (PPVs), negative predictive values (NPVs), and odds ratios (ORs) for PJI were calculated for the abnormal findings. Overall image impression for PJI was evaluated. χ2, Fisher exact, t, and Mann-Whitney U tests and receiver operating characteristic analysis were used. Interobserver agreement was assessed with κ statistics.
Results: Forty-three joints in 36 patients (mean ± SD age, 75.4 ± 8.8 years; 30 women; hip [n = 29], knee [n = 12], and elbow [n = 2]) were evaluated. Eighteen joints (42%) were clinically diagnosed as PJI. The findings suggesting PJI were capsular thickening (PPV, 70%; NPV, 90%; OR, 20.6), soft-tissue fluid collection (PPV, 81%; NPV, 81%; OR, 19.1), soft-tissue edema (PPV, 67%; NPV, 89%; OR, 17), pericapsular edema (PPV, 76%; NPV, 81%; OR, 13.7), and joint effusion (PPV, 55%; NPV, 100%; OR, 12). Soft-tissue fluid collection without capsular-like structure (PPV, 83%; NPV, 74%; OR, 14.4) or with communication to the joint (PPV, 75%; NPV, 71%; OR, 7.3) suggested PJI. The combinations of joint effusion, capsular thickening, pericapsular edema, soft-tissue fluid collection, and soft-tissue edema highly suggested PJI. Regarding the BME pattern, the combination with soft-tissue edema raised the possibility of PJI (PPV, 73%; NPV, 69%; OR, 5.9). Regarding the interobserver agreements for each abnormal finding, κ values were 0.60 to 0.77. Regarding the overall image impression, weighted κ value was 0.97 and areas under the receiver operating characteristic curve were 0.949 (95% confidence interval, 0.893-1.005) and 0.926 (95% confidence interval, 0.860-0.991) with no significant difference (P = 0.534).
Conclusions: The findings suggesting PJI were capsular thickening, soft-tissue fluid collection, soft-tissue edema, pericapsular edema, and joint effusion. The combinations of them highly suggested PJI. Regarding the BME pattern, the combination with soft-tissue edema raised the possibility of PJI.
Copyright © 2022 The Author(s). Published by Wolters Kluwer Health, Inc.
Conflict of interest statement
The authors declare no conflict of interest.
Similar articles
-
Evaluating Hip Periprosthetic Joint Infection with Metal-artifact-reduction MR Imaging.Magn Reson Med Sci. 2025 Jan 1;24(1):20-34. doi: 10.2463/mrms.mp.2023-0028. Epub 2024 Jan 16. Magn Reson Med Sci. 2025. PMID: 38233192 Free PMC article.
-
The role of advanced metal artifact reduction MRI in the diagnosis of periprosthetic joint infection.Skeletal Radiol. 2024 Oct;53(10):1969-1978. doi: 10.1007/s00256-023-04483-5. Epub 2023 Oct 24. Skeletal Radiol. 2024. PMID: 37875571 Free PMC article. Review.
-
Diagnosis of Periprosthetic Hip Joint Infection Using MRI with Metal Artifact Reduction at 1.5 T.Radiology. 2020 Jul;296(1):98-108. doi: 10.1148/radiol.2020191901. Epub 2020 May 12. Radiology. 2020. PMID: 32396046
-
Diagnostic accuracy of MRI with metal artifact reduction for the detection of periprosthetic joint infection and aseptic loosening of total hip arthroplasty.Eur J Radiol. 2020 Oct;131:109253. doi: 10.1016/j.ejrad.2020.109253. Epub 2020 Aug 31. Eur J Radiol. 2020. PMID: 32937252
-
Management of soft tissues in patients with periprosthetic joint infection.Arthroplasty. 2023 Oct 3;5(1):52. doi: 10.1186/s42836-023-00205-3. Arthroplasty. 2023. PMID: 37784182 Free PMC article. Review.
Cited by
-
Artificial Intelligence and Its Role in Predicting Periprosthetic Joint Infections.Biomedicines. 2025 Jul 30;13(8):1855. doi: 10.3390/biomedicines13081855. Biomedicines. 2025. PMID: 40868110 Free PMC article. Review.
-
Analysis of the clinical efficacy of two-stage revision surgery in the treatment of periprosthetic joint infection in the knee: A retrospective study.World J Clin Cases. 2022 Dec 26;10(36):13239-13249. doi: 10.12998/wjcc.v10.i36.13239. World J Clin Cases. 2022. PMID: 36683646 Free PMC article.
-
Evaluating Hip Periprosthetic Joint Infection with Metal-artifact-reduction MR Imaging.Magn Reson Med Sci. 2025 Jan 1;24(1):20-34. doi: 10.2463/mrms.mp.2023-0028. Epub 2024 Jan 16. Magn Reson Med Sci. 2025. PMID: 38233192 Free PMC article.
-
Imaging in Periprosthetic Joint Infection Diagnosis: A Comprehensive Review.Microorganisms. 2024 Dec 24;13(1):10. doi: 10.3390/microorganisms13010010. Microorganisms. 2024. PMID: 39858778 Free PMC article. Review.
-
The role of advanced metal artifact reduction MRI in the diagnosis of periprosthetic joint infection.Skeletal Radiol. 2024 Oct;53(10):1969-1978. doi: 10.1007/s00256-023-04483-5. Epub 2023 Oct 24. Skeletal Radiol. 2024. PMID: 37875571 Free PMC article. Review.
References
-
- Kurtz S, Ong K, Lau E, et al. Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am. 2007;89:780–785. doi:10.2106/JBJS.F.00222. - DOI
-
- Kapadia BH, Berg RA, Daley JA, et al. Periprosthetic joint infection. Lancet. 2016;387:386–394. doi:10.1016/S0140-6736(14)61798-0. - DOI
-
- Della Valle C, Parvizi J, Bauer TW, et al. The diagnosis of periprosthetic joint infections of the hip and knee. J Bone Joint Surg Am. 2011;93:1355–1357. doi:10.2106/JBJS.9314ebo. - DOI
-
- Fritz J, Lurie B, Miller TT, et al. MR imaging of hip arthroplasty implants. Radiographics. 2014;34:E106–E132. doi:10.1148/rg.344140010. - DOI
-
- Talbot BS, Weinberg EP. MR imaging with metal-suppression sequences for evaluation of total joint arthroplasty. Radiographics. 2016;36:209–225. doi:10.1148/rg.2016150075. - DOI
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous