Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2022 Jul;23(7):e13627.
doi: 10.1002/acm2.13627. Epub 2022 Apr 29.

Using 4D dose accumulation to calculate organ-at-risk dose deviations from motion-synchronized liver and lung tomotherapy treatments

Affiliations

Using 4D dose accumulation to calculate organ-at-risk dose deviations from motion-synchronized liver and lung tomotherapy treatments

William S Ferris et al. J Appl Clin Med Phys. 2022 Jul.

Abstract

Tracking systems such as Radixact Synchrony change the planned delivery of radiation during treatment to follow the target. This is typically achieved without considering the location changes of organs at risk (OARs). The goal of this work was to develop a novel 4D dose accumulation framework to quantify OAR dose deviations due to the motion and tracked treatment. The framework obtains deformation information and the target motion pattern from a four-dimensional computed tomography dataset. The helical tomotherapy treatment plan is split into 10 plans and motion correction is applied separately to the jaw pattern and multi-leaf collimator (MLC) sinogram for each phase based on the location of the target in each phase. Deformable image registration (DIR) is calculated from each phase to the references phase using a commercial algorithm, and doses are accumulated according to the DIR. The effect of motion synchronization on OAR dose was analyzed for five lung and five liver subjects by comparing planned versus synchrony-accumulated dose. The motion was compensated by an average of 1.6 cm of jaw sway and by an average of 5.7% of leaf openings modified, indicating that most of the motion compensation was from jaw sway and not MLC changes. OAR dose deviations as large as 19 Gy were observed, and for all 10 cases, dose deviations greater than 7 Gy were observed. Target dose remained relatively constant (D95% within 3 Gy), confirming that motion-synchronization achieved the goal of maintaining target dose. Dose deviations provided by the framework can be leveraged during the treatment planning process by identifying cases where OAR doses may change significantly from their planned values with respect to the critical constraints. The framework is specific to synchronized helical tomotherapy treatments, but the OAR dose deviations apply to any real-time tracking technique that does not consider location changes of OARs.

Keywords: Radixact; Synchrony; tomotherapy.

PubMed Disclaimer

Conflict of interest statement

John E. Bayouth has ownership interest in MR Guidance, LLC, which has business activity with a company that utilizes image guided radiation therapy technology (ViewRay, Inc.). Although this project was not sponsored externally, the data was collected on a Radixact system (Accuray Inc.) provided to UW–Madison under a research agreement (Bayouth, PI). Randall Kimple is supported in part by the University of Wisconsin Carbone Cancer Center Support Grant (No. P30 CA014520). The remaining authors have no conflicts of interest to disclose.

Figures

FIGURE 1
FIGURE 1
(a) Location of the target centroid in each phase relative to the reference phase derived from the 4DCT images for Lung 3. (b) The continuous respiratory amplitude pattern. (c) The discrete respiratory pattern mapped to each of the 10 phases. (d) The discrete mathematical target motion pattern. 4DCT, four‐dimensional computed tomography
FIGURE 2
FIGURE 2
Example of assigning sinogram projections from the original treatment plan to each phase of the 4DCT for Lung 3. The central 30 MLC leaves and the first 250 projections of the treatment are shown. 4DCT, four‐dimensional computed tomography; MLC, multi‐leaf collimator
FIGURE 3
FIGURE 3
Example of generation of motion‐synchronized treatment plans for Lung 3. (a) Differential sinograms expressed as the absolute difference between the original MLC sinogram and the motion‐synchronized sinogram. The central 30 MLC leaves and the first 250 projections of the treatment are shown. (b) Jaw field edges after compensation projected to isocenter for each phase as a function of projection in the treatment. The dotted gray lines indicate the centered jaw positions. MLC, multi‐leaf collimator
FIGURE 4
FIGURE 4
Dose differences (Synchrony accumulation minus plan) for each lung subject for the full prescription. For example, a 5‐Gy dose difference is a 10% difference relative to the prescription of 50 Gy
FIGURE 5
FIGURE 5
Dose differences (Synchrony accumulation minus plan) for each liver subject for the full prescription. For example, a 5‐Gy dose difference is a 10% difference relative to a prescription of 50 Gy

References

    1. Schnarr E, Beneke M, Casey D, et al. Feasibility of real‐time motion management with helical tomotherapy. Med Phys. 2018;45:1329‐1337. - PubMed
    1. Chao EH, Lucas D, Schnarr E. Evaluation of TomoTherapy dose calculations with intrafractional motion and motion compensation. Med Phys. 2018;45:18‐28. - PubMed
    1. Chen GP, Tai A, Keiper TD, Lim S, Li XA. Technical Note: comprehensive performance tests of the first clinical real‐time motion tracking and compensation system using MLC and jaws. Med Phys. 2020;47:2814‐2825. - PMC - PubMed
    1. Ferris WS, Kissick MW, Culberson WS, Smilowitz JB. Evaluation of Radixact motion synchrony for 3D respiratory motion: modeling accuracy and dosimetric fidelity. J Appl Clin Med Phys. 2020;21:96‐106. - PMC - PubMed
    1. Ferris WS, Culberson WS. Technical Note: patient dose from kilovoltage radiographs during motion‐synchronized treatments on Radixact. Med Phys. 2020;47:5772‐5778. - PMC - PubMed