Case Reports
. 2022 Apr 30;63(2):197-201.
doi: 10.3325/cmj.2022.63.197.

Senhance robot-assisted adrenalectomy: a case series

Affiliations
Case Reports

Senhance robot-assisted adrenalectomy: a case series

Nikola Knežević et al. Croat Med J. .

Abstract

We present a case series of 12 consecutive robot-assisted adrenalectomies performed from May 2019 to March 2020 by a single surgeon experienced in laparoscopy using the novel Senhance robotic system. Eleven patients had primary aldosteronism due to an adrenal adenoma, diagnosed by means of endocrinological and radiological evaluation, and 1 had a benign adrenal cyst. The robotic adrenalectomy technique is described in detail. The mean procedure time was 165.1 minutes, with robotic docking time of 11.6 minutes and console time of 98.6 minutes. The mean estimated blood loss was 47 mL, and hospital stay duration was 4.5 days. There was 1 Clavien Dindo IIIB complication and 1 patient underwent conversion to laparoscopy. All patients with adenoma had complete biochemical remission after surgery. In conclusion, the Senhance robotic system is a safe and feasible platform for benign adrenal surgery in high-volume centers.

PubMed Disclaimer

Figures

Figure 1
Figure 1
The difference between right side laparoscopic (blue circles) and robotic trocar position (red circles).

References

    1. Yiannakopoulou E. Robotic assisted adrenalectomy: Surgical techniques, feasibility, indications, oncological outcome and safety. Int J Surg. 2016;28:169–72. doi: 10.1016/j.ijsu.2016.02.089. - DOI - PubMed
    1. Samalavicius NE, Janusonis V, Siaulys R, Jasėnas M, Deduchovas O, Venckus R, et al. Robotic surgery using Senhance® robotic platform: single center experience with first 100 cases. J Robot Surg. 2020;14:371–6. doi: 10.1007/s11701-019-01000-6. - DOI - PubMed
    1. McCarus SD. Senhance robotic platform system for gynecological surgery. JSLS. 2021;25::e2020.00075. doi: 10.4293/JSLS.2020.00075. - DOI - PMC - PubMed
    1. Hirano Y, Kondo H, Miyawaki Y, Sugita H, Sakuramoto S, Yamaguchi S. Single-incision plus two-port robotic surgery for sigmoid colon cancer using the Senhance robotic system. Asian J Endosc Surg. 2021;14:94–6. doi: 10.1111/ases.12822. - DOI - PubMed
    1. Kastelan Z, Hudolin T, Kulis T, Knezevic N, Penezic L, Maric M, et al. Upper urinary tract surgery and radical prostatectomy with Senhance® robotic system: Single center experience-First 100 cases. Int J Med Robot. 2021;17:e2269. doi: 10.1002/rcs.2269. - DOI - PubMed

Publication types