Improving brain B0 shimming using an easy and accessible multi-coil shim array at ultra-high field
- PMID: 35511312
- PMCID: PMC9596507
- DOI: 10.1007/s10334-022-01014-6
Improving brain B0 shimming using an easy and accessible multi-coil shim array at ultra-high field
Abstract
Object: Improve shimming capabilities of ultra-high field systems, with addition of an accessible low-complexity B0 shim array for head MRI at 7 T.
Materials and methods: An eight channel B0 shim coil array was designed as a tradeoff between shimming improvement and construction complexity, to provide an easy to use shim array that can be employed with the standard 7 T head coil. The array was interfaced using an open-source eight-channel shim amplifier rack. Improvements in field homogeneity for whole-brain and slice-based shimming were compared to standard second-order shimming, and to more complex higher order dynamic shimming and shim arrays with 32 and 48 channels.
Results: The eight-channel shim array provided 12% improvement in whole brain static shimming and provided 33% improvement when using slice-based shimming. With this, the eight-channel array performed similar to third-order dynamic shimming (without the need for higher order eddy current compensation). More complex shim arrays with 32 and 48 channels performed better, but require a dedicated RF coil.
Discussion: The designed eight-channel shim array provides a low-complexity and low-cost approach for improving B0 field shimming on an ultra-high field system. In both static and dynamic shimming, it provides improved B0 homogeneity over standard shimming.
Keywords: B0 field; Magnetic resonance imaging; Multi-coil array; Shimming; Ultra-high field.