A Novel Dipotassium Hydrogen Phosphate Phantom for Calibrating Computed Tomographic Bone Density Measurements in Birds
- PMID: 35526162
- DOI: 10.1647/20-00018
A Novel Dipotassium Hydrogen Phosphate Phantom for Calibrating Computed Tomographic Bone Density Measurements in Birds
Abstract
The objective of this study was to construct a calibration phantom for bone mineral density (BMD) measurements adapted to avian anatomy by quantitative computed tomography. The determination of BMD is important to assess avian osteoporosis in poultry at production facilities and to study biological features in association with flight patterns in birds. Quantitative computed tomography measured in Hounsfield units is a well-established technique for BMD measurements. Translation of Hounsfield units into the International System of Units (mg/cm3) requires the use of a calibration phantom. Although calibration phantoms for routine use in humans are commercially available, phantoms suited to avian anatomy are not. A liquid dipotassium hydrogen phosphate calibration standard was constructed out of commercially available materials, easily allowing for variations in size, bone diameter, and adaptation to avian skeletal anatomy. Periodically, quantitative computed tomography scans were performed to monitor constant correlation to the calibration standard over 3 months and to monitor for the potential influence of gas bubbling and water evaporation in the rods on BMD measurements. Finally, the calibration phantom was tested for BMD measurements with carcasses from 2 bird species, including 3 peregrine falcons (Falco peregrinus; 2 juvenile males, 1 adult female with inactive reproductive status) and 4 Eurasian kestrels (Falco tinnunculus; 1 juvenile and 2 adult males, 1 adult female with inactive reproductive status). Results demonstrated stability of the calibration phantom without the need to refill or replace rods, plus a stable correlation line (R 2 = .99) over the 3-month evaluation period. It was possible to place the phantom directly on the bird carcasses, close to the measured bones, to improve BMD analysis. As evaluated, the phantom appeared to be adaptive to avian skeletal anatomy. Moreover, it was possible to build the phantom within 24 hours from commercially available materials.
Keywords: K2HPO4; avian; bone mineral density; dipotassium hydrogen phosphate; phantom; quantitative computed tomography; raptor.
Similar articles
-
Liquid Calibration Phantoms in Ultra-Low-Dose QCT for the Assessment of Bone Mineral Density.J Clin Densitom. 2020 Jan-Mar;23(1):108-116. doi: 10.1016/j.jocd.2019.02.005. Epub 2019 Feb 22. J Clin Densitom. 2020. PMID: 30902572
-
Improving Bone Mineral Density Assessment Using Spectral Detector CT.J Clin Densitom. 2019 Jul-Sep;22(3):374-381. doi: 10.1016/j.jocd.2018.10.004. Epub 2018 Oct 26. J Clin Densitom. 2019. PMID: 30497869
-
Quantitative micro-computed tomography: a non-invasive method to assess equivalent bone mineral density.Bone. 2008 Aug;43(2):302-311. doi: 10.1016/j.bone.2008.04.009. Epub 2008 Apr 30. Bone. 2008. PMID: 18539557
-
Application of quantitative computed tomography for assessment of trabecular bone mineral density, microarchitecture and mechanical property.Clin Imaging. 2016 Mar-Apr;40(2):330-8. doi: 10.1016/j.clinimag.2015.09.016. Epub 2015 Sep 21. Clin Imaging. 2016. PMID: 26602163 Review.
-
Quantitative Computed Tomography (QCT) derived Bone Mineral Density (BMD) in finite element studies: a review of the literature.J Exp Orthop. 2016 Dec;3(1):36. doi: 10.1186/s40634-016-0072-2. Epub 2016 Dec 9. J Exp Orthop. 2016. PMID: 27943224 Free PMC article. Review.
References
-
- Korver DR, Saunders-Blades JL, Nadeau KL. Assessing bone mineral density in vivo: quantitative computed tomography. Poult Sci. 2004;83(2):222–229.
-
- Dumont ER. Bone density and the lightweight skeletons of birds. Proc Biol Sci. 2010;277(1691):2193–2198.
-
- Kiang JH. Avian Wing Bones. San Diego, CA: ProQuest Dissertations Publishing; 2013.
-
- Hester PY, Schreiweis MA, Orban JI, et al. Assessing bone mineral density in vivo: dual energy x-ray absorptiometry. Poult Sci. 2004;83(2):215–221.
-
- Fleming RH, Korver D, McCormack HA, Whitehead CC. Assessing bone mineral density in vivo: digitized fluoroscopy and ultrasound. Poult Sci. 2004;83(2):207–214.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Research Materials