Liquid chromatographic determination of cyclosporine in whole blood with the advanced automated sample processing unit
- PMID: 3558587
- DOI: 10.1016/s0021-9673(01)94645-x
Liquid chromatographic determination of cyclosporine in whole blood with the advanced automated sample processing unit
Abstract
We describe a rapid, precise, cost-effective, and accurate isocratic liquid chromatographic (LC) procedure for determining cyclosporine in whole blood. The cyclosporine is extracted from 0.5 ml of whole blood together with 200 micrograms of cyclosporin D, added per liter as internal standard, by using an Advanced Automated Sample Processing (AASP) unit. The on-line solid-phase extraction is performed on an octasilane sorbent cartridge which is interfaced with a Perkin-Elmer 83 X 4.6 mm I.D. cartridge column, packed with 3-micron octadecyl packing. The column is eluted with a mobile phase containing acetonitrile-water (13:7) at a flow-rate of 1.0 ml/min at a column temperature of 70 degrees C. The column effluent is monitored at 210 nm. The absolute recovery of cyclosporine exceeded 87% and the linearity extended up to 2000 micrograms/l. Within-run and day-to-day coefficients of variation were less than 8%. The correlation between AASP-LC and manual Bond-Elut extraction-LC method was excellent (r = 0.97).
Similar articles
-
Automated solid-phase extraction and liquid chromatography for assay of cyclosporine in whole blood.Clin Chem. 1987 Dec;33(12):2272-4. Clin Chem. 1987. PMID: 3690843
-
Solid-phase extraction and liquid chromatography for improved assay of cyclosporine in whole blood or plasma.Clin Chem. 1985 Oct;31(10):1717-20. Clin Chem. 1985. PMID: 4042335
-
New automated high-performance liquid chromatographic analysis of cyclosporin A and G in human serum.J Chromatogr. 1987 Jan 23;413:131-40. doi: 10.1016/0378-4347(87)80220-7. J Chromatogr. 1987. PMID: 3558662
-
Micro method for liquid chromatographic determination of cyclosporin A in whole blood with use of a rapid extraction procedure.J Anal Toxicol. 1991 Mar-Apr;15(2):95-7. doi: 10.1093/jat/15.2.95. J Anal Toxicol. 1991. PMID: 2051752
-
Semi-automated high-performance liquid chromatographic determination of cyclosporine A in whole blood using one-step sample purification and column-switching.J Chromatogr. 1985 Jun 14;341(2):411-9. doi: 10.1016/s0378-4347(00)84054-2. J Chromatogr. 1985. PMID: 3897257