A 2D strip ionization chamber array with high spatiotemporal resolution for proton pencil beam scanning FLASH radiotherapy

Affiliations

A 2D strip ionization chamber array with high spatiotemporal resolution for proton pencil beam scanning FLASH radiotherapy

Yunjie Yang et al. Med Phys. 2022 Aug.

Abstract

Purpose: Experimental measurements of two-dimensional (2D) dose rate distributions in proton pencil beam scanning (PBS) FLASH radiation therapy (RT) are currently lacking. In this study, we characterize a newly designed 2D strip-segmented ionization chamber array (SICA) with high spatial and temporal resolution and demonstrate its applications in a modern proton PBS delivery system at both conventional and ultrahigh dose rates.

Methods: A dedicated research beamline of the Varian ProBeam system was employed to deliver a 250-MeV proton PBS beam with nozzle currents up to 215 nA. In the research and clinical beamlines, the spatial, temporal, and dosimetric performances of the SICA were characterized and compared with measurements using parallel-plate ion chambers (IBA PPC05 and PTW Advanced Markus chamber), a 2D scintillator camera (IBA Lynx), Gafchromic films (EBT-XD), and a Faraday cup. A novel reconstruction approach was proposed to enable the measurement of 2D dose and dose rate distributions using such a strip-type detector.

Results: The SICA demonstrated a position accuracy of 0.12 ± 0.02 mm at a 20-kHz sampling rate (50 μs per event) and a linearity of R2 > 0.99 for both dose and dose rate with nozzle beam currents ranging from 1 to 215 nA. The 2D dose comparison to the film measurement resulted in a gamma passing rate of 99.8% (2 mm/2%). A measurement-based proton PBS 2D FLASH dose rate distribution was compared to simulation results and showed a gamma passing rate of 97.3% (2 mm/2%).

Conclusions: The newly designed SICA demonstrated excellent spatial, temporal, and dosimetric performances and is well suited for commissioning, quality assurance, and a wide range of clinical applications in proton PBS clinical and FLASH-RT.

Keywords: FLASH-RT; dose rate monitoring; pencil beam scanning; proton therapy; strip ionization chamber; ultrahigh dose rate.

PubMed Disclaimer

References

REFERENCES

    1. Favaudon V, Caplier L, Monceau V, et al. Ultrahigh dose-rate FLASH irradiation increases the differential response between normal and tumor tissue in mice. Sci Transl Med. 2014;6(245):245ra93. https://doi.org/10.1126/scitranslmed.3008973
    1. Vozenin MC, Hendry JH, Limoli CL. Biological benefits of ultra-high dose rate FLASH radiotherapy: sleeping beauty awoken. Clin Oncol (R Coll Radiol). 2019;31(7):407-415. https://doi.org/10.1016/j.clon.2019.04.001
    1. Wilson JD, Hammond EM, Higgins GS, et al. Ultra-high dose rate (FLASH) radiotherapy: silver bullet or fool's gold? Front Oncol. 2020;9:1563. https://doi.org/10.3389/fonc.2019.01563
    1. Chow R, Kang M, Wei S, et al. FLASH radiation therapy: review of the literature and considerations for future research and proton therapy FLASH trials. Appl Radiat Oncol. 2021;10(2):16-21.
    1. Montay-Gruel P, Petersson K, Jaccard M, et al. Irradiation in a flash: unique sparing of memory in mice after whole brain irradiation with dose rates above 100Gy/s. Radiother Oncol. 2017;124(3):365-369. https://doi.org/10.1016/j.radonc.2017.05.003

LinkOut - more resources