Electronic scanning-slit fluorography: design and performance of a prototype unit
- PMID: 3561336
- DOI: 10.1118/1.596094
Electronic scanning-slit fluorography: design and performance of a prototype unit
Abstract
Electronic scanning-slit fluorography involves replacing paired fore and aft slits for scatter rejection with only one beam-defining tantalum fore aperture. Since the video signal within the projection of the aperture on the image intensifier is much more intense than behind the tantalum, one can discriminate electronically between these two signals and thus eliminate the unwanted x-ray scatter and veiling glare. The general features of a prototype unit are described along with the rationale for the choice of design factors employed. Imaging time of 1-2 s has been achieved using multiple scanning slits. Small focal-spot size and large number of pixels are favored for higher dose utilization, shorter imaging time, and lower x-ray tube loading, as well as for better spatial resolution. Images of a chest phantom show better visibility of low-contrast details, especially in poorly penetrated areas, when compared with the image obtained and displayed under the same conditions, but using a conventional grid to reject scattered radiation.
Similar articles
-
Electronic scanning-slit fluorography.Acta Radiol Suppl. 1989;373:1-48. Acta Radiol Suppl. 1989. PMID: 2488063 Review.
-
Scatter rejection by electronic collimation.Med Phys. 1986 Mar-Apr;13(2):158-63. doi: 10.1118/1.595940. Med Phys. 1986. PMID: 3702810
-
Comparison of scatter rejection and low-contrast performance of scan equalization digital radiography (SEDR), slot-scan digital radiography, and full-field digital radiography systems for chest phantom imaging.Med Phys. 2011 Jan;38(1):23-33. doi: 10.1118/1.3519903. Med Phys. 2011. PMID: 21361171 Free PMC article.
-
Digital radiographic imaging system with multiple-slit scanning x-ray beam: preliminary report.Radiology. 1986 Nov;161(2):513-8. doi: 10.1148/radiology.161.2.3763924. Radiology. 1986. PMID: 3763924
-
Contrast and scatter in x-ray imaging.Radiographics. 1991 Mar;11(2):307-23. doi: 10.1148/radiographics.11.2.2028065. Radiographics. 1991. PMID: 2028065 Review.
Cited by
-
Inverse geometry x-ray imaging: application in interventional procedures.J Am Coll Radiol. 2011 Jan;8(1):74-7. doi: 10.1016/j.jacr.2010.11.002. J Am Coll Radiol. 2011. PMID: 21211771 Free PMC article. No abstract available.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources