Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2022 Jun:96:105665.
doi: 10.1016/j.clinbiomech.2022.105665. Epub 2022 May 6.

Reverse total shoulder arthroplasty baseplate stability with locking vs. non-locking peripheral screws

Affiliations

Reverse total shoulder arthroplasty baseplate stability with locking vs. non-locking peripheral screws

Elise J Martin et al. Clin Biomech (Bristol). 2022 Jun.

Abstract

Background: There are many options for glenosphere baseplate fixation commercially available, yet there is little biomechanical evidence supporting one type of fixation over another. In this study, we compared the biomechanical fixation of a reverse total shoulder glenoid baseplate secured with locking or non-locking peripheral screws.

Methods: Both a non-augmented mini baseplate with full backing support and an augmented baseplate were testing after implantation in solid rigid polyurethane foam. Each baseplate was implanted with a 30 mm central compression screw and four peripheral screws, either locking or non-locking (15 mm anterior/posterior and 30 mm superior/inferior). A 1 Hz cyclic force of 0-750 N was applied at a 60o angle for 5000 cycles. Throughout the test, the displacement of the baseplate was measured using a 3D Digital Image Correlation System.

Findings: The amount of migration measured in the both the non-augmented and augment cases shows no significant differences between locking and non-locking cases at the final cycle count (non-augment: 5.66 +/- 2.29 μm vs. 3.71 +/- 1.23 μm; p = 0.095, augment: 15.43 +/- 8.49 μm vs. 12.46 +/- 3.24 μm; p = 0.314). Additionally, the amount of micromotion measured for both sample types shows the same lack of significant difference (non-augment: 10.79 +/- 5.22 μm vs. 10.16 +/- 7.61 μm; p = 0.388, augment: 55.03 +/- 10.13 μm vs. 54.84 +/- 10.65 μm; p = 0.968).

Interpretation: The presence of locking versus non-locking peripheral screws does not make a significant difference on the overall stability of a glenoid baseplate, in both a no defect case with a non-augmented baseplate and a bone defect case with an augmented baseplate.

Keywords: 3D digital image correlation; Biomechanics; Glenoid baseplate; Peripheral screw; Reverse total shoulder arthroplasty; Total shoulder arthroplasty.

PubMed Disclaimer

Publication types

LinkOut - more resources