Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2022 Sep:107:13-18.
doi: 10.1016/j.clinbiochem.2022.05.010. Epub 2022 Jun 1.

The development and validation of a high-capacity serological assay for celiac disease

Affiliations

The development and validation of a high-capacity serological assay for celiac disease

Rolf A Klaasen et al. Clin Biochem. 2022 Sep.

Abstract

Background: The aim of the present study was to develop and clinically validate a high-throughput assay for serum IgA and IgG antibodies against transglutaminase-2 (TG2) and to determine appropriate assay cut-offs for large-scale population screening for celiac disease.

Method: An automated method was developed using dual label time-resolved fluorometry on the AutoDELFIA platform. Individuals (n = 1920) from the general population were screened. Subjects with serum anti-TG2 concentrations above a preliminary cut-off (>0.3 mg*/L anti-TG2 IgA or >0.5 mg*/L anti-TG2 IgG) were offered endoscopic examination and biopsy. A diagnosis of celiac disease was given if villous atrophy (Marsh grade 3) was found.

Results: The assay had a limit of quantification of 0.25 mg*/L (anti-TG2 IgA) and 0.60 mg*/L (anti-TG2 IgG) with imprecision (CV) < 16% and <18% respectively. A total of 66 individuals were above the preliminary cut-off, and 56 underwent endoscopy. Of these, 26 were diagnosed with celiac disease. Sixty-eight percent of subjects with anti-TG2 IgA ≥ 0.7 mg*/L or anti-TG2 IgG ≥ 1.0 mg*/L had biopsy-proven celiac disease, and utilization of these higher cut-offs identified 96% of biopsy-positive patients. At the time of endoscopy, all individuals with anti-TG2 IgA > 2.0 mg*/L had celiac disease, and this cut-off identified 88% of newly diagnosed celiac patients. Eight percent (2/26) of the newly diagnosed patients had primarily anti-TG2 IgG.

Conclusions: In this study we developed and clinically validated a robust and automated assay suitable for celiac disease screening in the general population.

Keywords: Anti-TG2; Assay validation; Automation; Celiac disease; Population screening.

PubMed Disclaimer

LinkOut - more resources