. 2022 Jun 3:16:119-123.
doi: 10.1016/j.artd.2022.05.003. eCollection 2022 Aug.

Dissociation of Bipolar Hemiarthroplasty of the Hip and Review of Literature

Affiliations
Case Reports

Dissociation of Bipolar Hemiarthroplasty of the Hip and Review of Literature

Athanasiou Vasileios et al. Arthroplast Today. .

Abstract

Component dissociation after a bipolar hemiarthroplasty is an uncommon complication that usually necessitates reoperation due to difficulties in closed reduction. To our knowledge, only a few cases have been published in the recent literature. We present a case of disassembly location of in a 68-year-old woman who underwent bipolar hemiarthroplasty 10 years ago due to a left hip femoral neck fracture. She began complaining about pain and difficulty to bear weight after an unintentional internal rotation movement of her left hip. The initial radiographs showed the disassembly of the bipolar hip prosthesis, and the patient underwent conversion to total hip arthroplasty. We conducted a literature review to explain the potential causes and mechanisms.

Keywords: Bipolar hemiarthroplasty; Disassembly; Dissociation; Inner head displacement.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Patient’s initial postoperative anteroposterior radiograph of a left hip bipolar hemiarthroplasty.
Figure 2
Figure 2
Anteroposterior radiograph demonstrating a dissociation of a left hip bipolar hemiarthroplasty. The outer head is in extreme varus position (arrow)—osteoarthritic changes of the acetabulum.
Figure 3
Figure 3
Retrieved bipolar component and femoral head (a). No macroscopic wear was found at the articulating surface of the polyethylene insert (b). Excessive wear was found at the slots and groove of the polyethylene that contributed to the locking mechanism (c).
Figure 4
Figure 4
Postoperative anteroposterior radiograph after conversion of the bipolar hemiarthroplasty to total hip arthroplasty.
Figure 5
Figure 5
Anteroposterior radiograph of the left hip at 1-month follow-up after conversion to a total hip arthroplasty.

References

    1. Lestrange N.R. Bipolar arthroplasty for 496 hip fractures. Clin Orthop Relat Res. 1990;251:7–19. - PubMed
    1. LaBelle L.W., Colwill J.C., Swanson A.B. Bateman bipolar hip arthroplasty for femoral neck fractures. A five- to ten-year follow-up study. Clin Orthop Relat Res. 1990;251:20–25. - PubMed
    1. Calder S.J., Anderson G.H., Jagger C., Harper W.M., Gregg P.J. Unipolar or bipolar prosthesis for displaced intracapsular hip fracture in octogenarians: a randomised prospective study. J Bone Joint Surg Br. 1996;78:391–394. - PubMed
    1. Ng D.Z., Lee K.B. Unipolar versus bipolar hemiarthroplasty for displaced femoral neck fractures in the elderly: is there a difference? Ann Acad Med Singap. 2015;44:197–201. - PubMed
    1. Yang B., Lin X., Yin X.M., Wen X.Z. Bipolar versus unipolar hemiarthroplasty for displaced femoral neck fractures in the elder patient: a systematic review and meta-analysis of randomized trials. Eur J Orthop Surg Traumatol. 2015;25:425–433. - PubMed