Application of digital PCR to determine the reliability of Xpert Xpress SARS-CoV-2 assay with envelope (E) gene negative and nucleocapsid (N2) gene positive results
- PMID: 35691105
- PMCID: PMC9119709
- DOI: 10.1016/j.diagmicrobio.2022.115726
Application of digital PCR to determine the reliability of Xpert Xpress SARS-CoV-2 assay with envelope (E) gene negative and nucleocapsid (N2) gene positive results
Abstract
This study used digital polymerase chain reaction (dPCR) to determine whether envelope (E) gene-negative and nucleocapsid (N2) gene-positive (E-N+) results obtained with the Cepheid Xpert Xpress SARS-CoV-2 assay are reliable. Using droplet digital PCR results as a reference, 18 of 22 E-N+ samples with a low viral load (81.8%) were identified as true positives.
Keywords: Cepheid Xpert Xpress SARS-CoV-2 assay; E gene negative; N2 gene positive; SARS-CoV-2; digital PCR; low viral load.
Copyright © 2022 Elsevier Inc. All rights reserved.
Conflict of interest statement
Declaration of competing interest The authors declare that we have no conflicts of interest.
References
-
- Khoshchehreh M, Wald-Dickler N, Holtom P, Butler-Wu SM. A needle in the haystack? Assessing the significance of envelope (E) gene-negative, nucleocapsid (N2) gene-positive SARS-CoV-2 detection by the Cepheid Xpert Xpress SARS-COV-2 assay. J Clin Virol. 2020;133 doi: 10.1016/j.jcv.2020.104683. - DOI - PMC - PubMed
-