Ultrasound Transducer Pressure: An Unexplored Source of Abdominal Aortic Aneurysm Measurement Error

Affiliations
Free article

Ultrasound Transducer Pressure: An Unexplored Source of Abdominal Aortic Aneurysm Measurement Error

Qasam Mohammed Ghulam et al. Ultrasound Med Biol. 2022 Sep.
Free article

Abstract

This study was aimed at quantifying the abdominal aortic aneurysm (AAA) compression phenomenon and assessing the use of a new tool to estimate transducer pressure in vivo. In this cross-sectional study, 47 participants with AAA and a median anterior-to-posterior (AP) AAA diameter of 46 mm (range: 30-76 mm) were included. The majority of the patients were overweight with body mass indexes >25 (33/47, 70%). A standardized ultrasound (US)-compatible gel pad, with a pre-defined thickness (15 mm) and mass (150 g), was interposed between the US transducer and participant to estimate the applied transducer pressure. A firm transducer pressure significantly lowered the median AP diameter from 46.1 mm (range: 29.7-76.3) to 39.4 mm (range: 21.7-67.5) (p = 0.001). The mean diameter bias between light and firm transducer pressure was 6.1 mm (95% confidence interval: 4.9-7.3). The applied transducer pressure varied significantly and could be measured by translating the deformation of a low-tech gel pad interposed between the US transducer and the participant.

Keywords: Abdominal aortic aneurysm; Measurement; Pressure; Transducer; Ultrasound; error.