Passively scanned, single-fiber optical coherence tomography probes for gastrointestinal devices
- PMID: 35708124
- PMCID: PMC9541095
- DOI: 10.1002/lsm.23576
Passively scanned, single-fiber optical coherence tomography probes for gastrointestinal devices
Abstract
Background/objectives: Optical coherence tomography (OCT) uses low coherence interferometry to obtain depth-resolved tissue reflectivity profiles (M-mode) and transverse beam scanning to create images of two-dimensional tissue morphology (B-mode). Endoscopic OCT imaging probes typically employ proximal or distal mechanical beam scanning mechanisms that increase cost, complexity, and size. Here, we demonstrate in the gastrointestinal (GI) tracts of unsedated human patients, that a passive, single-fiber probe can be used to guide device placement, conduct device-tissue physical contact sensing, and obtain two-dimensional OCT images via M-to-B-mode conversion.
Materials and methods: We designed and developed ultrasmall, manually scannable, side- and forward-viewing single fiber-optic probes that can capture M-mode OCT data. Side-viewing M-mode OCT probes were incorporated into brush biopsy devices designed to harvest the microbiome and forward-viewing M-mode OCT probes were integrated into devices that measure intestinal potential difference (IPD). The M-mode OCT probe-coupled devices were utilized in the GI tract in six unsedated patients in vivo. M-mode data were converted into B-mode images using an M-to-B-mode conversion algorithm. The effectiveness of physical contact sensing by the M-mode OCT probes was assessed by comparing the variances of the IPD values when the probe was in physical contact with the tissue versus when it was not. The capacity of forward- and side-viewing M-mode OCT probes to produce high-quality B-mode images was compared by computing the percentages of the M-to-B-mode images that showed close contact between the probe and the luminal surface. Passively scanned M-to-B-mode images were qualitatively compared to B-mode images obtained by mechanical scanning OCT tethered capsule endomicroscopy (TCE) imaging devices.
Results: The incorporation of M-mode OCT probes in these nonendoscopic GI devices safely and effectively enabled M-mode OCT imaging, facilitating real-time device placement guidance and contact sensing in vivo. Results showed that M-mode OCT contact sensing improved the variance of IPD measurements threefold and side-viewing probes increased M-to-B-mode image visibility by 10%. Images of the esophagus, stomach, and duodenum generated by the passively scanned probes and M-to-B-mode conversion were qualitatively superior to B-mode images obtained by mechanically scanning OCT TCE devices.
Conclusion: These results show that passive, single optical fiber OCT probes can be effectively utilized for nonendoscopic device placement guidance, device contact sensing, and two-dimensional morphologic imaging in the human GI tract in vivo. Due to their small size, lower cost, and reduced complexity, these M-mode OCT probes may provide an easier avenue for the incorporation of OCT functionality into endoscopic/nonendoscopic devices.
Keywords: B-mode OCT; M-mode OCT; endoscopic probe.
© 2022 The Authors. Lasers in Surgery and Medicine published by Wiley Periodicals LLC.
Similar articles
-
Common-path optical coherence tomography with side-viewing bare fiber probe for endoscopic optical coherence tomography.Rev Sci Instrum. 2007 Nov;78(11):113102. doi: 10.1063/1.2804112. Rev Sci Instrum. 2007. PMID: 18052460
-
Feasibility and Safety of Tethered Capsule Endomicroscopy in Patients With Barrett's Esophagus in a Multi-Center Study.Clin Gastroenterol Hepatol. 2022 Apr;20(4):756-765.e3. doi: 10.1016/j.cgh.2021.02.008. Epub 2021 Feb 4. Clin Gastroenterol Hepatol. 2022. PMID: 33549871 Free PMC article.
-
Endoscopic optical coherence tomography with a flexible fiber bundle.J Biomed Opt. 2018 Jun;23(6):1-8. doi: 10.1117/1.JBO.23.6.066001. J Biomed Opt. 2018. PMID: 29900706
-
Optical coherence tomography in detection of dysplasia and cancer of the gastrointestinal tract and bilio-pancreatic ductal system.World J Gastroenterol. 2008 Nov 14;14(42):6444-52. doi: 10.3748/wjg.14.6444. World J Gastroenterol. 2008. PMID: 19030194 Free PMC article. Review.
-
Optical coherence tomography in gastroenterology: a review and future outlook.J Biomed Opt. 2017 Dec;22(12):1-17. doi: 10.1117/1.JBO.22.12.121716. J Biomed Opt. 2017. PMID: 29260538 Review.
Cited by
-
The LMIT: Light-mediated minimally-invasive theranostics in oncology.Theranostics. 2024 Jan 1;14(1):341-362. doi: 10.7150/thno.87783. eCollection 2024. Theranostics. 2024. PMID: 38164160 Free PMC article. Review.
-
Study on the application of optical coherence microscopy in Hirschsprung's disease.Sci Rep. 2023 Feb 6;13(1):2083. doi: 10.1038/s41598-023-28341-5. Sci Rep. 2023. PMID: 36746975 Free PMC article.
References
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources