Multicenter Study
. 2022 Jun 16;22(1):271.
doi: 10.1186/s12872-022-02708-4.

Short term outcome after left atrial appendage occlusion with the AMPLATZER Amulet and WATCHMAN device: results from the ORIGINAL registry (saxOnian RegIstry analyzinG and followINg left atrial Appendage cLosure)

Affiliations
Multicenter Study

Short term outcome after left atrial appendage occlusion with the AMPLATZER Amulet and WATCHMAN device: results from the ORIGINAL registry (saxOnian RegIstry analyzinG and followINg left atrial Appendage cLosure)

Lucie Kretzler et al. BMC Cardiovasc Disord. .

Abstract

Background: Various randomized multicenter studies have shown that percutaneous left atrial appendage closure (LAAC) is not inferior in stroke prevention compared to vitamin K antagonists (VKA) and can be performed safely and effectively.

Aims: The prospective multicenter ORIGINAL registry in the Free State of Saxony (saxOnian RegIstry analyzinG and followINg left atrial Appendage cLosure) investigated the efficiency and safety of LAAC with Watchman or Amulet device in a real word setting. A special focus was put on the influence of LAAC frequency on periprocedural efficiency and safety.

Methods and results: The total of 482 consecutive patients (Abbott Amulet N = 93 and Boston Scientific Watchman N = 389) were included in the periinterventional analyses. After 6 weeks, 353 patients completed the first follow-up including transoesophageal echocardiography (TEE) (73.2%). Successful LAAC could be performed in more than 94%. The complication rate does not significantly differ between device types (p = 0.92) according to Fischer test and comprised 2.2% in the Amulet and 2.3% in the Watchman group. The kind of device and the frequency of LAAC per study center had no influence on the success and complication rates. Device related thrombus could be revealed more frequently in the Watchman group (4.5%) than in the Amulet group (1.4%) but this difference is still not significant in Fisher test (p = 0.14). Same conclusion can be made about residual leakage 1.1% versus 0% [not significant in Fisher test (p = 0.26)]. Dual antiplatelet therapy followed the intervention in 64% and 22% of patients were discharged under a combination of an anticoagulant (VKA/DOAC/Heparin) and one antiplatelet agent.

Conclusions: The ORIGINAL registry supports the thesis from large, randomized trials that LAAC can be performed with a very high procedural success rate in the everyday clinical routine irrespective of the used LAA device (Watchman or Amulet). The postprocedural antithrombotic strategy differs widely among the participating centers. Trial registration Name of the registry: "saxOnian RegIstry analyzinG and followINg left atrial Appendage cLosure", Trial registration number: DRKS00023803; Date of registration: 15/12/2020 'Retrospectively registered'; URL of trial registry record: https://www.drks.de/drks_web/navigate.do?navigationId=trial.HTML&TRIAL_ID=DRKS00023803 .

Keywords: Atrial fibrillation; Left atrial appendage; Patient registry; Stroke prevention.

PubMed Disclaimer

Conflict of interest statement

The authors declare no competing interests.

None of the authors has any conflicts of interest or competing interests to declare for this specific project.

Figures

Fig. 1
Fig. 1
Periprocedural data during LAAC. A total procedure time in min; B implantation success in %; C number of discarded devices/LAAC procedure; D fluoroscopy time in minutes; E contrast agent consumption in ml. LAAC/year left atrial appendage closure procedure/year/study center
Fig. 2
Fig. 2
Anticoagulation drug regimens after interventional left atrial appendage closure. DAPT dual antiplatelet therapy, NOAC novel anticoagulant, VKA vitamin K antagonist

References

    1. Hindricks G, Potpara T, Dagres N, Arbelo E, Bax JJ, Blomstrom-Lundqvist C, Boriani G, Castella M, Dan GA, Dilaveris PE, et al. 2020 ESC guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS): The Task Force for the diagnosis and management of atrial fibrillation of the European Society of Cardiology (ESC) developed with the special contribution of the European Heart Rhythm Association (EHRA) of the ESC. Eur Heart J. 2021;42(5):373–498. doi: 10.1093/eurheartj/ehaa612. - DOI - PubMed
    1. Masoudi FA, Calkins H, Kavinsky CJ, Slotwiner DJ, Turi ZG, Drozda JP, Jr, Gainsley P, American College of Cardiology. Heart Rhythm Society. Society for Cardiovascular Angiography and Interventions et al. 2015 ACC/HRS/SCAI Left atrial appendage occlusion device societal overview: a professional societal overview from the American College of Cardiology, Heart Rhythm Society, and Society for Cardiovascular Angiography and Interventions. Catheter Cardiovasc Interv. 2015;86(5):791–807. doi: 10.1002/ccd.26170. - DOI - PubMed
    1. Holmes DR, Jr, Kar S, Price MJ, Whisenant B, Sievert H, Doshi SK, Huber K, Reddy VY. Prospective randomized evaluation of the Watchman Left Atrial Appendage Closure device in patients with atrial fibrillation versus long-term warfarin therapy: the PREVAIL trial. J Am Coll Cardiol. 2014;64(1):1–12. doi: 10.1016/j.jacc.2014.04.029. - DOI - PubMed
    1. Reddy VY, Doshi SK, Sievert H, Buchbinder M, Neuzil P, Huber K, Halperin JL, Holmes D. Percutaneous left atrial appendage closure for stroke prophylaxis in patients with atrial fibrillation: 2.3-year follow-up of the PROTECT AF (Watchman Left Atrial Appendage System for Embolic Protection in Patients with Atrial Fibrillation) Trial. Circulation. 2013;127(6):720–729. doi: 10.1161/CIRCULATIONAHA.112.114389. - DOI - PubMed
    1. Boersma LV, Ince H, Kische S, Pokushalov E, Schmitz T, Schmidt B, Gori T, Meincke F, Protopopov AV, Betts T, et al. Efficacy and safety of left atrial appendage closure with WATCHMAN in patients with or without contraindication to oral anticoagulation: 1-year follow-up outcome data of the EWOLUTION trial. Heart Rhythm. 2017;14(9):1302–1308. doi: 10.1016/j.hrthm.2017.05.038. - DOI - PubMed