Temperature-controlled high-power short-duration ablation with 90 W for 4 s: outcome, safety, biophysical characteristics and cranial MRI findings in patients undergoing pulmonary vein isolation
- PMID: 35748975
- DOI: 10.1007/s10840-022-01146-8
Temperature-controlled high-power short-duration ablation with 90 W for 4 s: outcome, safety, biophysical characteristics and cranial MRI findings in patients undergoing pulmonary vein isolation
Abstract
Background: High-power short-duration (HPSD) radiofrequency ablation (RFA) is highly efficient and safe while reducing procedure and RF time in pulmonary vein isolation (PVI). The QDot™ catheter is a novel contact force ablation catheter that allows automated flow and power adjustments depending on the local tissue temperature to maintain a target temperature during 90 W/4 s lesions. We analysed intraprocedural data and periprocedural safety using the QDot-catheter in patients undergoing PVI for paroxysmal atrial fibrillation (PAF).
Methods: We included n = 48 patients undergoing PVI with the QDot-catheter with a temperature-controlled HPSD ablation mode with 90 W/4 s (TC-HPSD). If focal reconnection occurred besides repeat ablation, the ablation mode was changed to 50 W/15 s (QMode). N = 23 patients underwent cerebral MRI to detect silent cerebral lesions.
Results: Mean RF time was 8.1 ± 2.8 min, and procedure duration was 84.5 ± 30 min. The overall maximal measured catheter tip temperature was 52.0 °C ± 4.6 °C, mean overall applied current was 871 mA ± 44 mA and overall applied energy was 316 J ± 47 J. The mean local impedance drop was 12.1 ± 2.4 Ohms. During adenosine challenge, n = 14 (29%) patients showed dormant conduction. A total of n = 24 steam pops were detected in n = 18 patients (39.1%), while no pericardial tamponade occurred. No periprocedural thromboembolic complications occurred, while n = 4 patients (17.4%) showed silent cerebral lesion.
Conclusions: TC-HPSD ablation with 90 W/4 s using the QDot-catheter led to a reduction of procedure and RF time, while no major complications occurred. Despite optimized temperature control and power adjustment, steam pops occurred in a rather high number of patients, while none of them leads to tamponade or to clinical or neurological deficits.
Keywords: Ablation; Atrial fibrillation; High-power short duration; RF ablation.
© 2022. The Author(s), under exclusive licence to Springer Science+Business Media, LLC, part of Springer Nature.
References
-
- Calkins H, Hindricks G, Cappato R, Kim YH, Saad EB, Aguinaga L, Akar JG, Badhwar V, Brugada J, Camm J, Chen PS, Chen SA, Chung MK, Nielsen JC, Curtis AB, Davies DW, Day JD, d’Avila A, de Groot N, Di Biase L, Duytschaever M, Edgerton JR, Ellenbogen KA, Ellinor PT, Ernst S, Fenelon G, Gerstenfeld EP, Haines DE, Haissaguerre M, Helm RH, Hylek E, Jackman WM, Jalife J, Kalman JM, Kautzner J, Kottkamp H, Kuck KH, Kumagai K, Lee R, Lewalter T, Lindsay BD, Macle L, Mansour M, Marchlinski FE, Michaud GF, Nakagawa H, Natale A, Nattel S, Okumura K, Packer D, Pokushalov E, Reynolds MR, Sanders P, Scanavacca M, Schilling R, Tondo C, Tsao HM, Verma A, Wilber DJ, Yamane T. 2017 HRS/EHRA/ECAS/APHRS/SOLAECE expert consensus statement on catheter and surgical ablation of atrial fibrillation. Heart Rhythm. 2017;14:e275–444. - DOI - PubMed - PMC
-
- Macle L, Khairy P, Weerasooriya R, Novak P, Verma A, Willems S, Arentz T, Deisenhofer I, Veenhuyzen G, Scavee C, Jais P, Puererfellner H, Levesque S, Andrade JG, Rivard L, Guerra PG, Dubuc M, Thibault B, Talajic M, Roy D, Nattel S. investigators At: Adenosine-guided pulmonary vein isolation for the treatment of paroxysmal atrial fibrillation: an international, multicentre, randomised superiority trial. Lancet. 2015;386:672–9. - DOI - PubMed
-
- Kottmaier M, Popa M, Bourier F, Reents T, Cifuentes J, Semmler V, et al. Safety and outcome of very high power short duration ablation using 70 Watts for pul- monary vein isolation in patients with paroxysmal atrial fibrillation. Europace. 2019. https://doi.org/10.1093/europace/euz342 . - DOI
MeSH terms