. 2022 Oct;44(5):854-860.
doi: 10.1111/ijlh.13920. Epub 2022 Jun 24.

The diagnostic accuracy of Sysmex XN for identification of pseudothrombocytopenia using various thresholds for definition of platelet aggregation

Affiliations

The diagnostic accuracy of Sysmex XN for identification of pseudothrombocytopenia using various thresholds for definition of platelet aggregation

Hanne Elisabeth Lunde et al. Int J Lab Hematol. 2022 Oct.

Abstract

Introduction: The aim of this study was to investigate the diagnostic accuracy of the flag PLT-Clumps from the WNR/WDF and the PLT-F channels from Sysmex XN and to study how different cut-offs for investigation for pseudothrombocytopenia (PTCP) and the definition of platelet aggregation affected the diagnostic accuracy.

Methods: A smear review was performed for samples with platelet count <150 × 109 /L and samples flagged for platelet aggregation by Sysmex XN-20. The samples were investigated for platelet aggregation in 30 fields using 40× objective. Findings were classified by size and quantity using two definitions of aggregation: the Norwegian quality improvement of laboratory investigations (Noklus) and the Groupe Francophone de'Hematologie Cellulaire (GFHC). The Q-values for the PLT-Clumps flag from the WNR/WFD channel and the PLT-F channel were compared with smear findings.

Results: ROC analysis showed that the diagnostic accuracy of the PLT-clumps flags increased with increasing stringency of the definition of platelet aggregation and when only samples with thrombocytopenia were investigated. With the most stringent PTCP definitions, the diagnostic accuracy of the PLT-Clumps flag from PLT-F was very high (AUC 0.97-0.98) and markedly better than for WNR/WDF.

Conclusion: The diagnostic accuracy of PLT-F from Sysmex XN-20 for identification of PTCP was very good and superior to the WNDR/WDF channel in samples with platelet count <150 × 109 /L and a moderate to high number of aggregates in the smear. There is a need for a more precise definition of platelet aggregation.

Keywords: Sysmex; hematology instrument flag; platelet aggregation; platelets; pseudothrombocytopenia.

PubMed Disclaimer

Conflict of interest statement

The authors have no competing interests.

Figures

FIGURE 1
FIGURE 1
(A) Inclusion of samples. (B) Graphic presentation of 30 fields in a smear investigated for platelet aggregation. Created in BioRender
FIGURE 2
FIGURE 2
Percentage difference between platelet count in Na‐Citrate and EDTA‐blood as a function of the observed number of aggregates in 30 fields with a 40× objective (n = 44)

Similar articles

Cited by

References

    1. Schuff‐Werner P, Mansour J, Gropp A. Pseudo‐thrombocytopenia (PTCP). A challenge in the daily laboratory routine? J Lab Med. 2020;44:000010151520200099. 10.1515/labmed-2020-0099 - DOI
    1. International Society for Laboratory Hematology (ISLH) . Consensus Guidelines: Rules Internett 07.09.2020, Updated 2020. Accessed September 7, 2020. http://www.islh.org/web/consensus_rules.php
    1. Baccini V, Geneviève F, Jacqmin H, et al. Platelet counting: ugly traps and good advice. Proposals from the French‐speaking cellular hematology group (GFHC). J Clin Med. 2020;9(3):808. doi: 10.3390/jcm9030808 - DOI - PMC - PubMed
    1. Norsk kvalitetsforbedring av laboratorieundersøkelser (NOKLUS) . Anbefalinger når bør det lages blodutstryk basert på celletellinger og eller flagg fra hematologiinstrumenter. NOKLUS, 2019. Accessed June 26, 2019. https://www.noklus.no/helsepersonell-sykehus-og-private-laboratorier/med...
    1. Sysmex Europe GmbH . XN‐Series Flagging Interpretation Guide, version 1.0. Sysmex Product Managment Haematology & Integrated Laboratory Solutions, 2018.