The diagnostic accuracy of Sysmex XN for identification of pseudothrombocytopenia using various thresholds for definition of platelet aggregation
- PMID: 35751410
- PMCID: PMC9545299
- DOI: 10.1111/ijlh.13920
The diagnostic accuracy of Sysmex XN for identification of pseudothrombocytopenia using various thresholds for definition of platelet aggregation
Abstract
Introduction: The aim of this study was to investigate the diagnostic accuracy of the flag PLT-Clumps from the WNR/WDF and the PLT-F channels from Sysmex XN and to study how different cut-offs for investigation for pseudothrombocytopenia (PTCP) and the definition of platelet aggregation affected the diagnostic accuracy.
Methods: A smear review was performed for samples with platelet count <150 × 109 /L and samples flagged for platelet aggregation by Sysmex XN-20. The samples were investigated for platelet aggregation in 30 fields using 40× objective. Findings were classified by size and quantity using two definitions of aggregation: the Norwegian quality improvement of laboratory investigations (Noklus) and the Groupe Francophone de'Hematologie Cellulaire (GFHC). The Q-values for the PLT-Clumps flag from the WNR/WFD channel and the PLT-F channel were compared with smear findings.
Results: ROC analysis showed that the diagnostic accuracy of the PLT-clumps flags increased with increasing stringency of the definition of platelet aggregation and when only samples with thrombocytopenia were investigated. With the most stringent PTCP definitions, the diagnostic accuracy of the PLT-Clumps flag from PLT-F was very high (AUC 0.97-0.98) and markedly better than for WNR/WDF.
Conclusion: The diagnostic accuracy of PLT-F from Sysmex XN-20 for identification of PTCP was very good and superior to the WNDR/WDF channel in samples with platelet count <150 × 109 /L and a moderate to high number of aggregates in the smear. There is a need for a more precise definition of platelet aggregation.
Keywords: Sysmex; hematology instrument flag; platelet aggregation; platelets; pseudothrombocytopenia.
© 2022 The Authors. International Journal of Laboratory Hematology published by John Wiley & Sons Ltd.
Conflict of interest statement
The authors have no competing interests.
Figures


Similar articles
-
Definition of Significant Platelet Clumping: Should We Review All Samples With a Platelet Clumping Flag From Automated Hematology Analyzer?Int J Lab Hematol. 2025 Feb;47(1):79-86. doi: 10.1111/ijlh.14387. Epub 2024 Oct 21. Int J Lab Hematol. 2025. PMID: 39434430 Free PMC article.
-
Effects of time, anticoagulant and detection channel on platelet count in ethylenediaminetetraacetic acid (EDTA)-dependent pseudothrombocytopenia.PeerJ. 2025 Mar 4;13:e19103. doi: 10.7717/peerj.19103. eCollection 2025. PeerJ. 2025. PMID: 40061233 Free PMC article.
-
A Case of Pseudothrombocytopenia due to Pre-Analytical Issues Revealed by the Presence of Fibrin in a Blood Film.Clin Lab. 2024 May 1;70(5). doi: 10.7754/Clin.Lab.2023.231129. Clin Lab. 2024. PMID: 38747921
-
National recommendations of the Croatian Chamber of Medical Biochemists and Working group for Laboratory hematology of the Croatian Society of Medical Biochemistry and Laboratory Medicine: Management of samples with suspected EDTA-induced pseudothrombocytopenia.Biochem Med (Zagreb). 2024 Oct 15;34(3):030504. doi: 10.11613/BM.2024.030504. Biochem Med (Zagreb). 2024. PMID: 39435172 Free PMC article. Review.
-
Pseudothrombocytopenia with multiple anticoagulant sample collection tubes.Interv Med Appl Sci. 2016 Dec;8(4):181-183. doi: 10.1556/1646.8.2016.4.4. Interv Med Appl Sci. 2016. PMID: 28180009 Free PMC article. Review.
Cited by
-
Definition of Significant Platelet Clumping: Should We Review All Samples With a Platelet Clumping Flag From Automated Hematology Analyzer?Int J Lab Hematol. 2025 Feb;47(1):79-86. doi: 10.1111/ijlh.14387. Epub 2024 Oct 21. Int J Lab Hematol. 2025. PMID: 39434430 Free PMC article.
-
Effects of time, anticoagulant and detection channel on platelet count in ethylenediaminetetraacetic acid (EDTA)-dependent pseudothrombocytopenia.PeerJ. 2025 Mar 4;13:e19103. doi: 10.7717/peerj.19103. eCollection 2025. PeerJ. 2025. PMID: 40061233 Free PMC article.
References
-
- Schuff‐Werner P, Mansour J, Gropp A. Pseudo‐thrombocytopenia (PTCP). A challenge in the daily laboratory routine? J Lab Med. 2020;44:000010151520200099. 10.1515/labmed-2020-0099 - DOI
-
- International Society for Laboratory Hematology (ISLH) . Consensus Guidelines: Rules Internett 07.09.2020, Updated 2020. Accessed September 7, 2020. http://www.islh.org/web/consensus_rules.php
-
- Norsk kvalitetsforbedring av laboratorieundersøkelser (NOKLUS) . Anbefalinger når bør det lages blodutstryk basert på celletellinger og eller flagg fra hematologiinstrumenter. NOKLUS, 2019. Accessed June 26, 2019. https://www.noklus.no/helsepersonell-sykehus-og-private-laboratorier/med...
-
- Sysmex Europe GmbH . XN‐Series Flagging Interpretation Guide, version 1.0. Sysmex Product Managment Haematology & Integrated Laboratory Solutions, 2018.
MeSH terms
LinkOut - more resources
Full Text Sources