Changes in the peripapillary vasculature and macular thickness after cataract surgery using two phacoemulsification systems with optical coherence tomography angiography

Affiliations

Changes in the peripapillary vasculature and macular thickness after cataract surgery using two phacoemulsification systems with optical coherence tomography angiography

Liang-Zhang Tan et al. Int J Ophthalmol. .

Abstract

Aim: To assess the changes in the peripapillary vasculature and macular thickness after cataract surgery using two phacoemulsification systems with optical coherence tomography angiography (OCTA).

Methods: Fifty-two eyes of 52 patients with age-related cataract were randomized into two groups for phacoemulsification: Infiniti group (26 patients) using the Infiniti phacoemulsification system with gravity-fluidics and Centurion group (26 patients) using the Centurion phacoemulsification system with active-fluidics. The peripapillary vessel density (PVD) and macular thickness were examined using OCTA at baseline and at 1d, 1 and 3mo after cataract surgery.

Results: In the Infiniti group, the PVD was significantly reduced at 1d after the cataract surgery (P<0.001). However, the retinal nerve fiber layer (RNFL) thickness showed no significant change during the follow-up. Change in PVD 1d postoperatively was significantly negatively correlated to the cumulative dissipated energy (CDE), estimated fluid usage (EFU), effective phacoemulsification time (EPT), intraocular pressure (IOP), and total operating time (TOT; P<0.05). The macular thickness was significantly increased in all regions after the cataract surgery (P<0.05). However, no significant changes were found in the macular vessel density (VD) during the follow-up (P>0.05). In the Centurion group, the VD and thickness in the optic papilla and macula did not significantly change in all regions during the follow-up (all P>0.05). The best-corrected visual acuity (BCVA) significantly improved in both groups postoperatively (P<0.001).

Conclusion: Using the Infiniti phacoemulsification system, OCTA provides a promising analysis of retinal vascular alterations, demonstrating a reduction of the PVD and an increase in the macular thickness. The Centurion phacoemulsification system can provide better retinal vasculature preservation during cataract surgery.

Keywords: cataract surgery; optical coherence tomography angiography; phacoemulsification system; retinal vasculature.

PubMed Disclaimer

Figures

Figure 1
Figure 1. A 4.5×4.5-mm2 image of optic disc angiogram.
Figure 2
Figure 2. A 3×3-mm2 image of superficial capillary plexus angiogram.
Figure 3
Figure 3. The IOP, EPT, EFU, CDE and TOT are negatively correlated with the change in PVD at 1d postoperatively
IOP: Intraocular pressure; EPT: Effective phacoemulsification time; EFU: Estimated fluid usage; CDE: Cumulative dissipated energy; TOT: Total operating time; PVD: Peripapillary vessel density.

Similar articles

Cited by

References

    1. Micieli JA, Arshinoff SA. Cataract surgery. CMAJ. 2011;183(14):1621. - PMC - PubMed
    1. Malik PK, Dewan TR, Patidar AK, Sain E. Effect of IOP based infusion system with and without balanced phaco tip on cumulative dissipated energy and estimated fluid usage in comparison to gravity fed infusion in torsional phacoemulsification. Eye Vis (Lond) 2017;4:22. - PMC - PubMed
    1. Vasavada V, Raj SM, Praveen MR, Vasavada AR, Henderson BA, Asnani PK. Real-time dynamic intraocular pressure fluctuations during microcoaxial phacoemulsification using different aspiration flow rates and their impact on early postoperative outcomes: a randomized clinical trial. J Refract Surg. 2014;30(8):534–540. - PubMed
    1. Chen M, Anderson E, Hill G, Chen JJ, Patrianakos T. Comparison of cumulative dissipated energy between the Infiniti and Centurion phacoemulsification systems. Clin Ophthalmol. 2015;9:1367–1372. - PMC - PubMed
    1. Hilton EJR, Hosking SL, Gherghel D, Embleton S, Cunliffe IA. Beneficial effects of small-incision cataract surgery in patients demonstrating reduced ocular blood flow characteristics. Eye (Lond) 2005;19(6):670–675. - PubMed

LinkOut - more resources