The impact of drug error reduction software on preventing harmful adverse drug events in England: a retrospective database study
- PMID: 35820711
- PMCID: PMC9277403
- DOI: 10.1136/bmjoq-2021-001708
The impact of drug error reduction software on preventing harmful adverse drug events in England: a retrospective database study
Abstract
Introduction: The use of intravenous administration systems with dose error reduction software (DERS) is advocated to mitigate avoidable medication harm. No large-scale analysis of UK data has been attempted. This retrospective descriptive study aimed to estimate the prevalence of hard limit events and to estimate the potential severity of DERS events.
Method: Twelve months of DERS data was obtained from two NHS trusts in England. Definitions for drug categories and clinical areas were standardised and an algorithm developed to extract hard maximum (HMX) events. Subject matter experts (SMEs) were asked to rate severity of all HMX events on a scale of 0 (no harm) to 10 (death). These were analysed by clinical area and drug category, per 1000 administrations.
Results: A total of 745 170 infusions were administered over 644 052 patient bed days (PBDs). 45% of these (338 263) were administered with DERS enabled. HMX event incidence across the whole dataset was 17.9/1000 administrations (95% CI 17.5 to 18.4); 9.4/1000 PBDs (95% CI 9.2 to 9.7). 6067 HMX events were identified. 4604 were <2-fold deviations and excluded. HMX were identified in all drug categories. The highest incidence was antibacterial drugs (2.21%; 95% CI 2.13 to 2.29). Of the 1415 HMX events reviewed by SMEs, 747 (52.6%) were low/no harm. Drugs with greatest potential harm were antiarrhythmics (21.8/1000 administrations; 95% CI 16.3 to 29.1), parenteral anticoagulants (24.16/1000 administrations; 95% CI 15.3 to 37.9) and antiepileptics (20.86/1000 administrations; 95% CI 16.4 to 26.5). DERS has prevented severe harm or death in 110 patients in these hospitals. Medical and paediatric areas had higher prevalence of potentially harmful HMX events, but these were probably related to profile design.
Conclusion: Compliance with DERS in this study was 45%. DERS events are common, but potential harm is rare. DERS events are not related to specific clinical areas. There are some issues with definition and design of drug profiles that may cause DERS events, thus future work should focus on implementation and data standardisation for future large-scale analysis.
Keywords: Adverse events, epidemiology and detection; Incident reporting; Medication safety; Patient safety.
© Author(s) (or their employer(s)) 2022. Re-use permitted under CC BY-NC. No commercial re-use. See rights and permissions. Published by BMJ.
Conflict of interest statement
Competing interests: WSG was paid consultancy fees for database management and analysis under contract with the University of Manchester. AS and AP have received speaking honoraria from BD. MR and EW received honoraria from the University of Manchester for their assistance with the data analysis and interpretation.
Similar articles
-
Intravenous infusion practices across England and their impact on patient safety: a mixed-methods observational study.Southampton (UK): NIHR Journals Library; 2020 Feb. Southampton (UK): NIHR Journals Library; 2020 Feb. PMID: 32040286 Free Books & Documents. Review.
-
Intravenous Administration Errors Intercepted by Smart Infusion Technology in an Adult Intensive Care Unit.J Patient Saf. 2021 Sep 1;17(6):430-436. doi: 10.1097/PTS.0000000000000374. J Patient Saf. 2021. PMID: 28368966
-
Impact of a Review of a Smart Infusion Pump Library in Hematology/Oncology: Tailoring Content to Meet Specialty Needs.J Patient Saf. 2022 Apr 1;18(3):e640-e644. doi: 10.1097/PTS.0000000000000907. J Patient Saf. 2022. PMID: 34508039 Review.
-
Optimizing the Use of Dose Error Reduction Software on Intravenous Infusion Pumps.Hosp Pediatr. 2024 Jun 1;14(6):448-454. doi: 10.1542/hpeds.2023-007385. Hosp Pediatr. 2024. PMID: 38716570
-
Pharmacy-driven performance improvement initiative to increase compliance with intravenous smart pump drug error reduction systems at a large urban academic medical center.Am J Health Syst Pharm. 2024 May 7;81(10):361-369. doi: 10.1093/ajhp/zxad309. Am J Health Syst Pharm. 2024. PMID: 38069664
Cited by
-
A review of human factors and infusion pumps: lessons for procurement.Front Digit Health. 2025 Feb 4;7:1425409. doi: 10.3389/fdgth.2025.1425409. eCollection 2025. Front Digit Health. 2025. PMID: 39981104 Free PMC article. Review.
-
Evaluating the Economic Impact of the PedAMINES App in Reducing Medication Errors in Pediatric Emergency Care: Cost-Effectiveness Analysis.J Med Internet Res. 2024 Oct 25;26:e52077. doi: 10.2196/52077. J Med Internet Res. 2024. PMID: 39454199 Free PMC article.