. 2022 Sep;53(3):374-383.
doi: 10.1016/j.jmir.2022.06.009. Epub 2022 Jul 16.

Optimal theranostic SPECT imaging protocol for 223radium dichloride therapy

Affiliations

Optimal theranostic SPECT imaging protocol for 223radium dichloride therapy

Luis Felipe C Lima et al. J Med Imaging Radiat Sci. 2022 Sep.

Abstract

223Radium dichloride image-based individual dosimetry requires an optimal acquisition and reconstruction protocol and proper image correction methods for theranostic applications. To assess this problem, radium-223 dichloride SPECT images were acquired from a Jaszczak simulator with a dual-headed gamma camera, LEHR collimator, 128 × 128 matrix, and total time of 32 minutes. A cylindrical PMMA phantom was used to calibrate the measurements performed with Jaszczak. The image quality parameters (noise coefficient, contrast, contrast-to-noise ratio and recovery coefficient) and septal penetration correction were calculated by MATLAB®. The best results for the investigated image quality parameters were obtained with an 89 keV energy window (24% wide) produced together with OSEM/MLEM reconstruction (8 subsets and 4 iterations) applying a Butterworth filter (order 10 and cutoff frequency of 0.48 cycles·cm-1). The successfully performed recovery coefficient parameter evaluation allows uptake correction for future patient dosimetry applications.

PubMed Disclaimer

LinkOut - more resources