Comparative Study
. 1987 Jun;79(6):907-14.

Accuracy of blood pressure measurement by the Dinamap monitor in infants and children

  • PMID: 3588145
Comparative Study

Accuracy of blood pressure measurement by the Dinamap monitor in infants and children

M K Park et al. Pediatrics. 1987 Jun.

Erratum in

  • Pediatrics 1988 May;81(5):683

Abstract

Indirect BPs measured by the Dinamap monitor, an oscillometric device, and the conventional auscultatory method were compared with the direct radial artery pressure in infants and children. There were 29 patients in the Dinamap group with a median age of 18 months (ranges 1 month to 16 years) and 20 patients in the auscultatory group with a median age of 3.5 years (ranges 3 months to 16 years). The direct radial artery pressures were recorded on a strip chart and the ranges of pressures were obtained for systolic, diastolic, and mean pressures. The range of the direct readings was converted to a weighted single reading, and comparisons were made between the direct and indirect readings. The linear regression equation between the direct (x) and the Dinamap (y) readings (with correlation coefficient) was y = 1.05x - 5.36 (r = .97) for systolic, y = 1.10x - 4.65 (r = .903) for diastolic, and y = 1.06x - 4.21 (r = .917) for mean pressures. The linear regression equation (with correlation coefficient) for the auscultatory pressure was y = 1.60x - 68.23 (r = .872) and y = 1.38x - 16.47 (r = .874) for diastolic pressures. The error was defined as the indirect reading minus the direct reading. The mean error +/- SD was -0.24 +/- 3.26 for systolic, 1.28 +/- 4.74 for diastolic, and 0.10 +/- 4.56 mm Hg for mean pressures by the Dinamap method. It was -1.65 +/- 6.68 for systolic, and 8.70 +/- 5.97 for diastolic pressures by the auscultatory method.(ABSTRACT TRUNCATED AT 250 WORDS)