A Novel Tube Insertion Technique for Glaucoma Drainage Device Implantation
- PMID: 35882037
- DOI: 10.1097/IJG.0000000000002080
A Novel Tube Insertion Technique for Glaucoma Drainage Device Implantation
Abstract
Purpose: Early hypotony after non-valved glaucoma drainage device (GDD) implantation for complex glaucomatous eyes with labile aqueous production can lead to significant visual morbidity. We therefore sought to report the early postoperative outcomes of a novel surgical technique that allows atraumatic insertion of non-valved GDDs through a much smaller 25-gauge scleral track, to minimize entry site leakage and improve safety.
Methods: Retrospective case series of 15 consecutive cases undergoing non-valved GDD insertion into the anterior chamber using a previously unreported technique.
Results: All eyes underwent successful GDD insertion using our novel technique, with no intraoperative complications. The mean preoperative intraocular pressures (IOP), at day 1, week 1 and week 3 were 31.4, 22.4, 23.7, and 25.6 mm Hg, respectively. A statistically significant IOP reduction was achieved at day 1, week 1 and week 3 postoperatively ( P <0.05) without any observed leakage at the scleral entry site. One eye (6.7%) with complex panuveitic glaucoma developed early hypotony (5 mm Hg) with shallow choroidal detachments on day 1. This was successfully managed with 1 intracameral ophthalmic viscoelastic device injection given at the slit-lamp and no further intervention.
Conclusions: This novel single needle-docking intraocular insertion manoeuvre is an easily adoptable technique to make GDD insertion through a smaller 25-gauge water-tight scleral track more efficient and less traumatic. The technique reduces scleral distortion and therefore improves surgical safety particularly in eyes with complex secondary glaucoma.
Copyright © 2022 Wolters Kluwer Health, Inc. All rights reserved.
Conflict of interest statement
Disclosure: The authors declare no conflict of interest.
References
-
- Molteno CB, Polkinghorne J, Bowbyes. A. The vicryl tie technique for inserting a draining implant in the treatment of secondary glaucoma. Aust NZ J Ophthalmol. 1986;14:343–354.
-
- Egbert PR, Liebert MF. Internal suture occlusion of the Molteno glaucoma implant for the prevention of post-operative hypotony. Opthalmic Surg. 1989;20:53–56.
-
- Siegner SW, Netland PA, Urban RC Jr, et al. Clinical experience with Baerveldt glaucoma drainage implant. Ophthalmology. 1995;102:1298–1307.
-
- Daniel E, Pistilli M, Pujari SS, et al. Risk of hypotony in non-infectious uveitis. Ophthalmology. 2012;119:2377–2385.
-
- Lim KS, Garg A, Cheng J, et al. Comparison of short-term post-operative hypotony rates of 23-gauge vs 25-gauge needles in formation of the scleral tract for Baerveldt Tube insertion into the anterior chamber. J Curr Glaucoma Pract. 2018;12:36–39.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical