Local gentamicin-collagen sponge reduces cardiovascular implantable electronic device infections and pocket hematoma
- PMID: 35916820
- DOI: 10.26355/eurrev_202207_29311
Local gentamicin-collagen sponge reduces cardiovascular implantable electronic device infections and pocket hematoma
Abstract
Objective: Implantation or replacement of a cardiovascular implantable electronic device (CIED) may be associated with complications, such as pocket hematomas and infections. This study aims to determine whether a lyophilized gentamycin-containing collagen implant (GCCI) reduces major CIED infections and pocket hematomas after implant.
Subjects and methods: A retrospective study was conducted among patients who underwent implantation or replacement of CIED at the Tor Vergata Polyclinic (Rome, Italy) between June 2007 and November 2019. The primary combined endpoint was infection and hematoma occurrence through 12 months of follow-up post-procedure. The rate of single infectious complications, pocket hematomas or both were also assessed.
Results: We compared 475 patients treated with the GCCI (GCCI group) with 714 patients who did not receive it (control group). Complications occurred in 127 patients (11%); a statistically significant reduction of infections and pocket hematomas in the GCCI group was reported when compared with control patients (1% vs. 17%; p<0.0001). A total of 20 (2%) infectious events were reported, 102 (8%) patients developed a pocket hematoma, and 5 (0.4%) had both. The rate of single complications was significantly lower in GCCI group: infection 0.2% vs. 2.6% (p=0.002), pocket hematoma 0.6% vs. 13.8% (p<0.001). The association between antiplatelet/anticoagulation therapy and hematoma development was not statistically significant.
Conclusions: The GCCI is a medical device that can be used in addition to local hemostasis and prophylactic doses of systemic antibiotics with the aim of reducing infective complications and pocket hematoma after permanent CIED implantation or replacement.
Similar articles
-
Pocket related complications following cardiac electronic device implantation in patients receiving anticoagulation and/or dual antiplatelet therapy: prospective evaluation of different preventive strategies.J Interv Card Electrophysiol. 2019 Apr;54(3):247-255. doi: 10.1007/s10840-018-0488-y. Epub 2018 Nov 21. J Interv Card Electrophysiol. 2019. PMID: 30460588
-
Device implantation for patients on antiplatelets and anticoagulants: Use of suction drain.Indian Heart J. 2018 Dec;70 Suppl 3(Suppl 3):S389-S393. doi: 10.1016/j.ihj.2017.12.009. Epub 2018 Jan 2. Indian Heart J. 2018. PMID: 30595295 Free PMC article.
-
Use of a pocket compression device for the prevention and treatment of pocket hematoma after pacemaker and defibrillator implantation (STOP-HEMATOMA-I).J Interv Card Electrophysiol. 2017 Aug;49(2):197-204. doi: 10.1007/s10840-017-0235-9. Epub 2017 Apr 17. J Interv Card Electrophysiol. 2017. PMID: 28413854
-
Current strategies to minimize postoperative hematoma formation in patients undergoing cardiac implantable electronic device implantation: A review.Heart Rhythm. 2021 Apr;18(4):641-650. doi: 10.1016/j.hrthm.2020.11.017. Epub 2020 Nov 23. Heart Rhythm. 2021. PMID: 33242669 Review.
-
The relationship between pocket hematoma and risk of wound infection among patients with a cardiovascular implantable electronic device: An integrative review.Heart Lung. 2020 Jan-Feb;49(1):92-98. doi: 10.1016/j.hrtlng.2019.09.009. Epub 2019 Oct 10. Heart Lung. 2020. PMID: 31607409 Review.
Cited by
-
Reconstruction of Orbital Wall Fractures with a Combination of Resorbable Plates and Antibiotic-Impregnated Collagen Sheets.J Clin Med. 2024 Mar 25;13(7):1900. doi: 10.3390/jcm13071900. J Clin Med. 2024. PMID: 38610668 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical