Case Reports
. 2022 Sep-Oct;22(5):261-263.
doi: 10.1016/j.ipej.2022.07.008. Epub 2022 Aug 5.

Rise in ICD pacing impedance: What is the cause?

Affiliations
Case Reports

Rise in ICD pacing impedance: What is the cause?

Vickram Vignesh Rangaswamy et al. Indian Pacing Electrophysiol J. 2022 Sep-Oct.

Erratum in

No abstract available

PubMed Disclaimer

Conflict of interest statement

Declaration of competing interest None.

Figures

Fig. 1
Fig. 1
A) Graph showing right ventricular (RV) impedance trend over 80 week period. B) Real-time tracing at 25mm/s, showing EGMs. The source of EGM1, EGM2 (wavelet), EGM3 in the tracings are RV tip to RV ring, can to RV coil, and RVtip to RVcoil, respectively. (EGM–electrogram, Vs – sensed ventricular complex, Vp-paced ventricular complex, RV – right ventricle).
Fig. 2
Fig. 2
A) On interrogation, displaying live telemetry of sensed events (abbreviations as in Fig. 1 legend). B) On interrogation, EGMs recorded during ventricular pacing (abbreviation as in Fig. 1).

References

    1. Swerdlow C.D., Koneru J.N., Gunderson B., Kroll M.W., Ploux S., Ellenbogen K.A. Impedance in the diagnosis of lead malfunction. Circ Arrhythm Electrophysiol. 2020;13(2) - PubMed
    1. Mourad A.R., Kim M.H., Kirk M.M. A case series of anomalous high pacing lead impedances in normally functioning leads. Heart Rhythm Case Rep. 2015;1(6):449–452. - PMC - PubMed
    1. Hauser R.G., Sengupta J., Casey S., Tang C., Stanberry L.I., Abdelhadi R. High shocking and pacing impedances due to defibrillation lead calcification. J Intervent Card Electrophysiol. 2020;58(3):253–259. - PMC - PubMed
    1. Katsaras D., Cassidy C., Lakha M., Gordon D., Goode G., Abozguia K. Intermittent loss of telemetry data. JACC: Case Rep. 2021;3(1):146–149. - PMC - PubMed

Publication types