Sources of error in acid-base analysis from a blood gas analyser result: a narrative review
- PMID: 35934908
- DOI: 10.36303/JSAVA.163
Sources of error in acid-base analysis from a blood gas analyser result: a narrative review
Abstract
Preservation of blood pH within a narrow range is essential to optimal physiological function. This narrow pH range is maintained via the interactions of various buffer systems. Blood gas analysis is thus essential in the diagnosis and management of disorders affecting blood pH. Common methods of acid-base interpretation in veterinary science are the traditional approach, the physicochemical approach and the semiquantitative approach. However, blood gas analysis is prone to error during the preanalytical, analytical and post-analytical phases of the laboratory process. The pre-analytical phase incorporates steps in obtaining the sample, thus sources of pre-analytical error are related to operator technique. Most errors occur during the pre-analytical phase. Pre-analytical errors include entrainment of air bubbles into the sample and delays between sampling and analysis, both of which cause inaccurate measurement of oxygen and carbon dioxide tensions. The analytical phase outlines processes within the analyser. Common analytical errors are related to substances confounding analyte measurements. The post-analytical phase mainly describes interpretation of the results. Some of the approaches to acid-base interpretation require extensive post-analytical calculations, thus lending themselves to error. Errors occurring during the prior phases will be amplified. Errors in the measurement of the carbon dioxide tension (from which bicarbonate concentration and base excess are calculated) will introduce error into all three methods of acid-base interpretation. Furthermore, errors occurring in the measurements of electrolytes and lactate will result in incorrect interpretations if the physicochemical and semiquantitative approaches are applied. The potential sources of error during the various phases are reviewed.
Similar articles
-
Potential preanalytical errors in whole-blood analysis: effect of syringe sample volume on blood gas, electrolyte and lactate values.Scand J Clin Lab Invest. 2009;69(5):585-91. doi: 10.1080/00365510902878716. Scand J Clin Lab Invest. 2009. PMID: 19396657
-
[Sources of error in the pre-analytical phase of blood gas analysis].Praxis (Bern 1994). 1995 Feb 14;84(7):200-3. Praxis (Bern 1994). 1995. PMID: 7871307 German.
-
Determination of reference intervals for equine arterial blood-gas, acid-base and electrolyte analysis.Vet Anaesth Analg. 2019 Nov;46(6):765-771. doi: 10.1016/j.vaa.2019.04.015. Epub 2019 Jun 14. Vet Anaesth Analg. 2019. PMID: 31471125
-
The Pre-Analytical Aspects of Blood Gas Analysis Point-Of-Care-Testing (POCT) and Their Impact on Clinical Decision-Making.Clin Lab. 2025 May 1;71(5). doi: 10.7754/Clin.Lab.2024.241029. Clin Lab. 2025. PMID: 40387753 Review.
-
Pitfalls in the diagnosis and management of acid-base disorders in humans: a laboratory medicine perspective.J Clin Pathol. 2024 Oct 20;77(11):772-778. doi: 10.1136/jcp-2024-209423. J Clin Pathol. 2024. PMID: 39025490 Review.
Cited by
-
Impact of an air bubble within the syringe on test results obtained with a modern blood gas analyzer.J Med Biochem. 2024 Sep 6;43(5):690-695. doi: 10.5937/jomb0-49870. J Med Biochem. 2024. PMID: 39712506 Free PMC article.
-
Relationships between arterial and central venous blood acid-base variables in anaesthetised cats during euvolaemic and hypovolaemic states.J Feline Med Surg. 2025 Mar;27(3):1098612X241309829. doi: 10.1177/1098612X241309829. Epub 2025 Mar 22. J Feline Med Surg. 2025. PMID: 40119721 Free PMC article.