Methods for measuring pre-, intra-, and postoperative skin thickness for cochlear implants
- PMID: 35949548
- PMCID: PMC9349019
- DOI: 10.1016/j.joto.2021.12.001
Methods for measuring pre-, intra-, and postoperative skin thickness for cochlear implants
Abstract
Objective: This study was conducted to determine whether there is a reliable method for measuring the thickness of the retroauricular skin before, during, and after cochlear implantation, which allows the assessment of the optimal force of the external magnet of the cochlear implant (CI).
Methods: The retroauricular skin thickness of 83 patients who received a CI was measured using three different methods. The thickness was measured on pre- and postoperative CT images, as well as intraoperatively. The magnet category chosen by the surgeon was recorded when the implant was switched on and during the first follow-up visit. Correlation analyses were performed on the different skin thickness measurements and between the skin thickness and magnet strength categories.
Results: Only six patients required an exchange of the magnet until the follow-up. Although the median absolute thickness differed significantly between the three measures (p < 0.0001), their thickness values showed highly significant correlations (Pearson's r = 0.457-0.585; p < 0.01). In addition, magnet strength, was significantly correlated with the flap thickness determined pre-, post-, and during surgery. The lowest correlation with magnet strength was found in the intraoperative needle method.
Conclusion: All three measurements methods provided a suitable base for determining the ideal magnetic force. However, of particular interest were the pre- and postoperative CT measurements. The first enabled the early assessment of the required magnetic strength and thus a timely postoperative supply, whereas the latter helped to estimate the need for magnetic strength reduction during follow-up care and the feasibility of an early swith-on.
Keywords: Cochlear implant; Flap thickness; Headpiece; Magnet strength.
© 2021 PLA General Hospital Department of Otolaryngology Head and Neck Surgery. Production and hosting by Elsevier (Singapore) Pte Ltd.
Conflict of interest statement
The Author(s) declare(s) that there is no conflict of interest.
Figures




Similar articles
-
Predicting the Magnet Strength Using a Tablet-Based Tool for Patients Undergoing Cochlear Implantation.Cureus. 2023 Oct 3;15(10):e46417. doi: 10.7759/cureus.46417. eCollection 2023 Oct. Cureus. 2023. PMID: 37794885 Free PMC article.
-
Skin flap thickness and magnet strength in Cochlear implants.Cochlear Implants Int. 2021 Mar;22(2):80-84. doi: 10.1080/14670100.2020.1824640. Epub 2020 Sep 30. Cochlear Implants Int. 2021. PMID: 32993474
-
Blinded comparison of computed tomography, ultrasound and needle methods to measure skin flap thickness for cochlear implantation.Cochlear Implants Int. 2023 Nov;24(6):295-300. doi: 10.1080/14670100.2023.2239515. Epub 2023 Aug 23. Cochlear Implants Int. 2023. PMID: 37612887
-
First MRI With New Cochlear Implant With Rotatable Internal Magnet System and Proposal for Standardization of Reporting Magnet-Related Artifact Size.Otol Neurotol. 2019 Aug;40(7):883-891. doi: 10.1097/MAO.0000000000002269. Otol Neurotol. 2019. PMID: 31219967 Review.
-
Case report: traumatic displacement of a cochlear implant magnet.Ann Otol Rhinol Laryngol. 2014 Apr;123(4):229-31. doi: 10.1177/0003489414525922. Ann Otol Rhinol Laryngol. 2014. PMID: 24671476 Review.
Cited by
-
Triamcinolone Injection for Cochlear Implant Magnet Adherence Issues.Otol Neurotol. 2023 Sep 1;44(8):e572-e576. doi: 10.1097/MAO.0000000000003974. Epub 2023 Aug 3. Otol Neurotol. 2023. PMID: 37550870 Free PMC article.
References