Minute Stroke Distance Is a More Reproducible Measurement Than Cardiac Output in the Assessment of Fetal Ventricular Systolic Function

Affiliations

Minute Stroke Distance Is a More Reproducible Measurement Than Cardiac Output in the Assessment of Fetal Ventricular Systolic Function

Heather Y Sun et al. J Am Soc Echocardiogr. 2023 Feb.

Abstract

Background: Echocardiographic quantification of fetal cardiac output (CO) aids clinical decision-making in the management of various cardiac and extracardiac diseases. Small variability in measuring semilunar valve dimension significantly reduces the reproducibility of the calculated CO. The authors propose minute stroke distance or velocity-time integral (VTI) as a more reproducible measure reflecting fetal ventricular systolic function. The aim of this study was to test the hypothesis that right and left ventricular minute VTI increase predictably with estimated fetal weight and are more reproducible than CO.

Methods: Five hundred seventy-one singleton fetuses without cardiovascular pathology between 16 and 36 weeks' gestation were reviewed retrospectively. Twenty-two fetuses with pathology resulting in low- or high-CO states were also assessed for comparison. VTI was measured in both ventricular outflow tracts at the level of the semilunar valve, excluding a Doppler insonation angle of >30°. Heart rate, semilunar valve dimension, and VTI determined minute VTI and CO. Inter- and intrarater variability were evaluated in a random 10% subset.

Results: Minute VTI and CO measurements were feasible in 67% to 89% of fetuses in this retrospective study. Minute VTI and CO increased with estimated fetal weight nonlinearly (R = 0.61-0.94). The mean inter- and intrarater variability for VTI, 6% and 5.7%, were significantly less than for CO, 25% and 23.7% (P < .001 for all).

Conclusions: Minute VTI is an easily measured, highly reproducible method of quantifying fetal ventricular systolic function. Variability in calculated CO from valve measurement differences is minimized by solely using VTI. Nomograms of minute VTI provide an efficient and precise assessment of fetal systolic function and may be used to track fetuses in disease states with low or high CO.

Keywords: Cardiac output; Fetal echocardiography; Fetal ultrasonography; Stroke distance.