Trans-arterial Radioembolization Dosimetry in 2022
- PMID: 35982334
- DOI: 10.1007/s00270-022-03215-x
Trans-arterial Radioembolization Dosimetry in 2022
Abstract
Trans-arterial radioembolization is currently performed using 90Y-loaded glass or resin microspheres and also using 166Ho-loaded microspheres. The goal of this review is to present dosimetry and radiobiology concepts, the different dosimetry approaches available (simulation-based dosimetry and post-treatment dosimetry), main confounding factors as main clinical dosimetry results provided during the last decade for both hepatocellular carcinoma (HCC) and metastases of colorectal carcinoma (mCRC). Based on the different number of microspheres or different isotope used, radiobiology of the three devices is different, meaning that tumouricidal doses and maximal tolerated doses are different. Tumouricidal doses described for HCCs were 100-120 grays (Gy) with 90Y resin microspheres and 205 Gy with 90Y glass microspheres. For mCRC, it is 39-60 with 90Y resin microspheres, 139 Gy with 90Y glass microspheres and 90 Gy with 166Ho microspheres. An impact of tumoural doses with overall survival has also been reported. Personalised dosimetry has been developed and is now recommended by several international expert groups. Level-one evidence of the major impact of personalised dosimetry on response and overall survival in HCC is now available, bringing a new standard approach for TARE in clinical practice as well as for trial design.
Keywords: Dosimetry; Hepatocellular carcinoma; Radioembolization; mCRC.
© 2022. Springer Science+Business Media, LLC, part of Springer Nature and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE).
References
-
- Garin E, Rolland Y, Laffont S, Edeline J. Clinical impact of (99m)Tc-MAA SPECT/CT-based dosimetry in the radioembolization of liver malignancies with (90)Y-loaded microspheres. Eur J Nucl Med Mol Imag. 2016;43:559–75.
-
- Braat AJAT, Prince JF, van Rooij R, Bruijnen RCG, van den Bosch MAAJ, Lam MGEH. Safety analysis of holmium-166 microsphere scout dose imaging during radioembolisation work-up: a cohort study. Eur Radiol. 2018;28(3):920–8. - PubMed
-
- SIR-Spheres® Y-90 resin microspheres [package insert]. St Leonards, New South Wales, Australia: Sirtex Medical Pty Ltd; December 2019. https://www.sirtex.com/au/clinicians/package-insert/
-
- TheraSphere™ Yttrium-90 Glass Microspheres [package insert]. Marlborough, MA; Boston Scientific Corporation; 1999. https://btgplc.com/BTG/media/TheraSphere-Documents/PDF/10093509-Rev8_Eng...
Publication types
MeSH terms
Substances
LinkOut - more resources