Comparison of performance of self-expanding and balloon-expandable transcatheter aortic valves
- PMID: 36004225
- PMCID: PMC9390782
- DOI: 10.1016/j.xjon.2022.04.015
Comparison of performance of self-expanding and balloon-expandable transcatheter aortic valves
Abstract
Objective: To evaluate the flow dynamics of self-expanding and balloon-expandable transcatheter aortic valves pertaining to turbulence and pressure recovery. Transcatheter aortic valves are characterized by different designs that have different valve performance and outcomes.
Methods: Assessment of transcatheter aortic valves was performed using self-expanding devices (26-mm Evolut [Medtronic], 23-mm Allegra [New Valve Technologies], and small Acurate neo [Boston Scientific]) and a balloon-expandable device (23-mm Sapien 3 [Edwards Lifesciences]). Particle image velocimetry assessed the flow downstream. A Millar catheter was used for pressure recovery calculation. Velocity, Reynolds shear stresses, viscous shear stress, and pressure gradients were calculated.
Results: The maximal velocity at peak systole obtained with the Evolut R, Sapien 3, Acurate neo, and Allegra was 2.12 ± 0.19 m/sec, 2.41 ± 0.06 m/sec, 2.99 ± 0.10 m/sec, and 2.45 ± 0.08 m/sec, respectively (P < .001). Leaflet oscillations with the flow were clear with the Evolut R and Acurate neo. The Allegra shows the minimal range of Reynolds shear stress magnitudes (up to 320 Pa), and Sapien 3 the maximal (up to 650 Pa). The Evolut had the smallest viscous shear stress magnitude range (up to 3.5 Pa), and the Sapien 3 the largest (up to 6.2 Pa). The largest pressure drop at the vena contracta occurred with the Acurate neo transcatheter aortic valve with a pressure gradient of 13.96 ± 1.35 mm Hg. In the recovery zone, the smallest pressure gradient was obtained with the Allegra (3.32 ± 0.94 mm Hg).
Conclusions: Flow dynamics downstream of different transcatheter aortic valves vary significantly depending on the valve type, despite not having a general trend depending on whether or not valves are self-expanding or balloon-expandable. Deployment design did not have an influence on flow dynamics.
Keywords: PDF, probability density function; PG, pressure gradient; RSS, Reynolds shear stress; TAV, transcatheter aortic valve; TAVR; VSS, viscous shear stress; blood damage; pressure recovery; turbulence.
© 2022 The Author(s).
Figures
References
-
- Bui H.T., Khair N., Yeats B., Gooden S., James S.P., Dasi L.P. Transcatheter heart valves: a biomaterials perspective. Adv Healthc Mater. 2021;10:e2100115. - PubMed
-
- Abdel-Wahab M., Landt M., Neumann F.-J., Massberg S., Frerker C., Kurz T., et al. 5-Year outcomes after TAVR with balloon-expandable versus self-expanding valves: results from the CHOICE randomized clinical trial. Cardiovasc Interven. 2020;13:1071–1082. - PubMed
-
- Van Belle E., Vincent F., Labreuche J., Auffret V., Debry N., Lefèvre T., et al. Balloon-expandable versus self-expanding transcatheter aortic valve replacement: a propensity-matched comparison from the FRANCE-TAVI registry. Circulation. 2020;141:243–259. - PubMed
LinkOut - more resources
Full Text Sources